Moderate riskFDA DeviceClass II

Getinge SERVO Y sensor module recalled over missing safety tests

Getinge's Maquet Critical Care AB recalled nine Y Sensor modules (Model 6693662) for SERVO-U and SERVO-N ventilator systems. The modules did not complete electrical safety and functional tests during production. They were distributed to facilities in California, Georgia, Minnesota, and New York, and no injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Maquet recalls

What to do

  1. Check whether your facility uses a SERVO-U or SERVO-N ventilator with a Y Sensor module, Model 6693662.
  2. Compare the module's serial number to the recalled list: 21049, 21088, 21089, 21093, 21100, 21103, 21108, 21114, 21115.
  3. If your module is on the list, contact Getinge or Maquet Critical Care AB for instructions on next steps.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.
Sold in
Nationwide
Made for
All ages

Details

Units affected
9 units
Illnesses reported
None reported
Manufactured in
Rontgenvagen 2, Solna, Sweden
Classification
Class II
Recall number
Z-2963-2026

In the FDA's words

Description
GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge. The module did not undergo electrical safety tests and functional tests during production. DUI/DI 07325710000304, Serial/Lot Numbers: 21049, 21088, 21089, 21093, 21100, 21103 21108, 21114, 21115. Distribution: US Nationwide distribution in the states of CA, GA, MN, NY. Affected quantity: 9 units
Hazard
The module did not undergo electrical safety tests and functional tests during production.

Related topics

Electrical shockMedical devicesSold nationwideMore from FDA Device