IMRIS ORT200 operating room table recalled for mechanical lock failure risk
A mechanical failure in the linear actuator and rotational lock mechanism on certain IMRIS ORT200 operating room tables may prevent the table from locking or unlocking properly. If the table becomes unlocked unintentionally during surgery, it could move unexpectedly and injure the patient.
From the FDA notice · Read the official notice ↗ · All IMRIS recalls
What to do
- If your facility operates an affected ORT200 table (Part Numbers 113821-600 or 114148-000 with the listed serial numbers), stop using it immediately.
- Contact IMRIS Imaging Inc for instructions on inspection, repair, or replacement.
- Check the serial number on your table against the list in the FDA notice to confirm if your unit is affected.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 2 products — see all of them.
Details
- Units affected
- 6 units
- Illnesses reported
- None reported
- Manufactured in
- 1230 Chaska Creek Way Ste 100, N/A, Chaska, MN, United States
- Classification
- Class II
- Recall number
- Z-3033-2026
In the FDA's words
- Description
- ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000 The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure. 113821-600 - Serial Numbers: 10003754, 10006997; 114148-000 - UDI-DI: 00857534006349, Serial Numbers: 10005189, 10005190, 10004463, 10004997 Distribution: Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden. Affected quantity: 6 units
- Hazard
- The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
2 products in this recall
IMRIS ORT200 and ORT300 operating tables recalled for locking risk
IMRIS is recalling 2 operating room table models, the ORT200 and ORT300, with 19 affected units. A mechanical failure in the linear actuator and rotational lock can stop the table from switching between locked and unlocked, which could allow unintended rotation during a procedure. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2… (this page)Z-3033-2026 · 6 units
- ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600Z-3034-2026 · 13 units