Serious riskFDA DeviceClass II

IMRIS ORT200 operating room table recalled for mechanical lock failure risk

A mechanical failure in the linear actuator and rotational lock mechanism on certain IMRIS ORT200 operating room tables may prevent the table from locking or unlocking properly. If the table becomes unlocked unintentionally during surgery, it could move unexpectedly and injure the patient.

From the FDA notice · Read the official notice ↗ · All IMRIS recalls

What to do

  1. If your facility operates an affected ORT200 table (Part Numbers 113821-600 or 114148-000 with the listed serial numbers), stop using it immediately.
  2. Contact IMRIS Imaging Inc for instructions on inspection, repair, or replacement.
  3. Check the serial number on your table against the list in the FDA notice to confirm if your unit is affected.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 2 products — see all of them.

Details

Units affected
6 units
Illnesses reported
None reported
Manufactured in
1230 Chaska Creek Way Ste 100, N/A, Chaska, MN, United States
Classification
Class II
Recall number
Z-3033-2026

In the FDA's words

Description
ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000 The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure. 113821-600 - Serial Numbers: 10003754, 10006997; 114148-000 - UDI-DI: 00857534006349, Serial Numbers: 10005189, 10005190, 10004463, 10004997 Distribution: Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden. Affected quantity: 6 units
Hazard
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

2 products in this recall

IMRIS ORT200 and ORT300 operating tables recalled for locking risk

IMRIS is recalling 2 operating room table models, the ORT200 and ORT300, with 19 affected units. A mechanical failure in the linear actuator and rotational lock can stop the table from switching between locked and unlocked, which could allow unintended rotation during a procedure. No injuries have been reported.

The agency lists each product separately. Check every one you might have.