Serious riskFDA DeviceClass II

Medline Open Heart Convenience Kit recalled for sterile barrier breach risk

Medline's Open Heart Convenience Kit (DYNJ43008N) contains a coronary cannula with a potential sterile barrier breach. If a compromised pouch seal isn't caught before use, the non-sterile device could cause infection or serious complications including blood cell breakdown, blood clots, or organ damage.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check your kit for lot numbers 25HLA462, 25HLA916, 25JLA847, 25LLA298, or 26CLA372.
  2. Inspect the pouch seal carefully before use.
  3. Do not use the kit if the seal appears compromised.
  4. Contact Medline or your hospital's medical device team for instructions.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
Sold in
Nationwide
Made for
Adults (0 months and older)

Details

Units affected
135 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, N/A, Northfield, IL, United States
Classification
Class II
Recall number
Z-3137-2026

In the FDA's words

Description
Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction. UDI-DI: 10198459445941 (each), 40198459445942 (case); Lot Numbers: 25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372 Distribution: US Nationwide distribution in the state of CA. Affected quantity: 135 units
Hazard
The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.

More from Medline

See all

Related topics

SterilityMedical devicesSold nationwideMore from FDA Device