Moderate riskFDA DeviceClass II

Datex-Ohmeda Aisys CS2 anesthesia system recalled for anesthetic delivery interruption

Certain Datex-Ohmeda Aisys CS2 anesthesia systems with specific electronic gas mixer assemblies can automatically switch to a backup oxygen-only mode. When this happens, the delivery of anesthetic medication stops, and the system triggers alarms. A clinician must manually intervene to restore anesthetic delivery, or the patient may not receive adequate anesthesia.

From the FDA notice · Read the official notice ↗

What to do

  1. Check if your serial number matches the affected units: ANAS01357, ANAP00609, ANAQ00988, ANAS00586, ANAR00543, ANAP01088, ANAM00443.
  2. If your system is affected, contact Datex-Ohmeda, Inc. for guidance on the electronic gas mixer assembly.
  3. Follow your facility's procedures for equipment review and clinical safety protocols.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 4 products — see all of them.

Details

Units affected
7 units
Illnesses reported
None reported
Manufactured in
3030 Ohmeda Dr, N/A, Madison, WI, United States
Classification
Class II
Recall number
Z-3191-2026

In the FDA's words

Description
Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000 Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 7 units
Hazard
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

4 products in this recall

Datex-Ohmeda anesthesia systems recalled for interrupted anesthetic delivery

Datex-Ohmeda is recalling four anesthesia system products, with 1,032 units affected. Certain gas mixer assemblies can automatically switch the system to Alternate O2 control, which delivers 100% oxygen and interrupts anesthetic agent delivery until a clinician intervenes. If that does not happen, the patient may experience decreased anesthetic effect. No injuries have been reported.

The agency lists each product separately. Check every one you might have.