Datex-Ohmeda Aisys CS2 anesthesia system recalled for anesthetic delivery interruption
Certain Datex-Ohmeda Aisys CS2 anesthesia systems with specific electronic gas mixer assemblies can automatically switch to a backup oxygen-only mode. When this happens, the delivery of anesthetic medication stops, and the system triggers alarms. A clinician must manually intervene to restore anesthetic delivery, or the patient may not receive adequate anesthesia.
From the FDA notice · Read the official notice ↗
What to do
- Check if your serial number matches the affected units: ANAS01357, ANAP00609, ANAQ00988, ANAS00586, ANAR00543, ANAP01088, ANAM00443.
- If your system is affected, contact Datex-Ohmeda, Inc. for guidance on the electronic gas mixer assembly.
- Follow your facility's procedures for equipment review and clinical safety protocols.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 4 products — see all of them.
Details
- Units affected
- 7 units
- Illnesses reported
- None reported
- Manufactured in
- 3030 Ohmeda Dr, N/A, Madison, WI, United States
- Classification
- Class II
- Recall number
- Z-3191-2026
In the FDA's words
- Description
- Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000 Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect. Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 7 units
- Hazard
- Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
4 products in this recall
Datex-Ohmeda anesthesia systems recalled for interrupted anesthetic delivery
Datex-Ohmeda is recalling four anesthesia system products, with 1,032 units affected. Certain gas mixer assemblies can automatically switch the system to Alternate O2 control, which delivers 100% oxygen and interrupts anesthetic agent delivery until a clinician intervenes. If that does not happen, the patient may experience decreased anesthetic effect. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000Z-3189-2026 · 475 units
- Aisys CS2, Datex-Ohmeda anesthesia system, Part Number 1011-9055-000Z-3190-2026 · 140 units
- Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000 (this page)Z-3191-2026 · 7 units
- Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU)…Z-3192-2026 · 410 units