Moderate riskFDA DeviceClass II

Heidelberg Engineering SPECTRALIS imaging platform recalled for loose wheels and tip-over risk

The wheels on Heidelberg Engineering's SPECTRALIS imaging device may come loose when the module is pushed or pulled during movement or transportation. If wheels loosen, the stand can become unstable and tip over, potentially causing impact or crushing injuries to nearby people.

From the FDA notice · Read the official notice ↗

What to do

  1. Stop using the device immediately if you notice any loose or unstable wheels.
  2. Contact Heidelberg Engineering for inspection and repair of your device.
  3. Keep the area around the device clear of people while it remains in use.
  4. Check the lot code on your device against the affected list provided by the manufacturer.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye
Sold in
Nationwide
Made for
All ages (under 100 years)

Details

Units affected
87
Illnesses reported
None reported
Manufactured in
Max-Jarecki-Str. 8, Heidelberg, Germany
Classification
Class II
Recall number
Z-3215-2026

In the FDA's words

Description
Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons. UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611 Distribution: Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam. Affected quantity: 87
Hazard
Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.