Olympus EZDilate Endoscopic Balloon Dilator recalled for incorrect inflation chart label
Some Olympus EZDilate Endoscopic Balloon Dilator devices (Model BD-420X-1355) have an incorrect label on the shaft. The inflation chart label may show balloon sizes of 18, 19, or 20 mm when the actual device is 11, 12, or 13 mm. This mislabeling could lead to confusion during use.
From the FDA notice · Read the official notice ↗ · All Olympus recalls
What to do
- If you use or stock this device (Model BD-420X-1355, Lot 460898), verify the actual balloon size before use.
- Contact Olympus Corporation of the Americas for replacement or further guidance.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 305 units
- Illnesses reported
- None reported
- Manufactured in
- 3500 Corporate Pkwy, Center Valley, PA, United States
- Classification
- Class II
- Recall number
- Z-3216-2026
In the FDA's words
- Description
- Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355. Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm, Model Number: BD-420X-1355. UDI-DI: 00821925042261. Lot Number: 460898. Distribution: International distribution in the country of Japan. Affected quantity: 305 units
- Hazard
- Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
More from Olympus
See all- Olympus Tripp-Lite P580-025 DisplayPort Cable recalled for video dropout risk
ModerateFDAOct 7, 2026
- Olympus BiCOAG Hemostasis Probe recalled for ceramic tip detachment risk
ModerateFDASep 30, 2026
- Olympus POWERSEAL 5mm curved jaw sealer recalled for incorrect device length labeling
ModerateFDASep 2, 2026