FDA-Device2026-10-07Class IIPROCESSING DEFECT

Cardinal Health Presource surgical kits recalled for fluid leakage risk

OTHERNationwide distributionMedical devices

What this means

Cardinal Health Presource Packs containing Stryker StrykeFlow PS and AHTO tube sets may leak fluid during use. Fluid can pool on the floor and create a slip and fall hazard for operating room staff.

  • Check your inventory against the affected catalog numbers (PB23GLNLAC, PB23GLNLA1–4, PB23GRNLD, PB23GRNLE, PB23GRNL5, PB23GRNL6, SOTDKGREFB) and lot numbers listed in the recall notice.
  • Do not use affected kits until you receive guidance from Cardinal Health.
  • Contact Cardinal Health customer service for instructions on return or replacement.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.

Intended for
Adults
Illnesses reported
None reported
Sold states
US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
Affected count
2957 kits
Manufactured in
3651 Birchwood Dr, Waukegan, IL, United States
Products
Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLAC; 2) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLA1; 3) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLA2; 4) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLA3; 5) KIT, GYN LAPAROSCOPY NEW HAN , Catalog Number: PB23GLNLA4; 6) KIT, GYN ROBOTIC NEW HAN , Catalog Number: PB23GRNLD; 7) KIT, GYN ROBOTIC NEW HAN , Catalog Number: PB23GRNLE; 8) KIT, GYN ROBOTIC NEW HAN , Catalog Number: PB23GRNL5; 9) KIT, GYN ROBOTIC NEW HAN , Catalog Number: PB23GRNL6; 10) MVHS GU GYN ROBOT PACK , Catalog Number: SOTDKGREFB

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3297-2026

More Cardinal Health recalls

Want us to check the next one for you?

We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Free · One email a week · Unsubscribe anytime
By signing up you agree to our Terms and Privacy Policy.