- 2025-07-02FDA-DeviceClass IIGE Healthcare Versana Ultrasound Systems, sold under the following names and Model/Catalog Numbers: 1) Versana Premier R3 VS, Model/Catalog Number 5938558; 2) Versana Premier R3 VA, Model/Catalog Number 5938559; 3) Versana Premier Lotus 4PP, Model/Catalog Number 5938560; 4) Versana Premier Lotus 5PP, Model/Catalog Number 5938561; 5) Versana Premier R3 VA Vet, Model/Catalog Number 5946387; 6) Versana Premier R3 VS Vet, Model/Catalog Number 5946481. 7) Versana Premier R3 Expert, Mode
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
- 2025-07-02FDA-DeviceClass IIGE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 750c A2, Model/REF Number 1012-9755-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 750 A2, Model/REF Number 1012-9750-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 650C A2, Model/REF Number 1012-9655-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare Carestation 650 SE, Model/REF Number 1012-9650-012
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 650 A2, Model/REF Number 1012-9650-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2025-05-07FDA-DeviceClass IGE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000
GE HealthCare has become aware that certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
- 2023-10-11FDA-DeviceClass IIF2-01 Frame, physiological patient monitor (with arrhythmia detection or alarms)
There is a potential interruption of data communication between E-modules inserted in the F2-01 Frame and CARESCAPE ONE and CARESCAPE Canvas 1000 patient monitors if the F2-01 Frame has not been powered down within the last 120 days.