- 2024-05-22FDA-DeviceClass IIThe Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VTT
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
- 2024-05-22FDA-DeviceClass IIThe Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VT
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
Get notified about new Sophysa S.A. recalls
Free weekly digest. We email when a new federal recall affects Sophysa S.A. or anything else in your household.
Get the Sunday Brief