- 2025-04-02FDA-DeviceCovidienClass IShiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
- 2025-03-26FDA-DeviceFUJIFILM Healthcare Americas CorporationClass IISynapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
The incorrect computed patient age is showing in VX for patients less than 3 months old.
- 2025-03-26FDA-DeviceAmerican Contract Systems IncClass IICRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.
During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.
- 2025-03-26FDA-DeviceSmiths Medical ASD, Inc.Class IICAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
- 2025-03-26FDA-DeviceSmiths Medical ASD, Inc.Class IICAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW1100
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
- 2025-03-26FDA-DeviceZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAYClass IIComprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
- 2025-03-26FDA-DeviceMerge Healthcare, Inc.Class IIMerge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-26FDA-DeviceMerge Healthcare, Inc.Class IIVERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-26FDA-DeviceMerge Healthcare, Inc.Class IIVERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-26FDA-DeviceMerge Healthcare, Inc.Class IIVERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-26FDA-DeviceMerge Healthcare, Inc.Class IIVERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- 2025-03-26FDA-DeviceB Braun Medical IncClass IIIntrocan Safety 3 Closed IV Catheter 22 Ga. x 1 in., PUR, Winged. Passive safety IV catheter.
Increased risk for valve housing detachment causing leakage.
- 2025-03-26FDA-DeviceRemel, IncClass IIZebra Thermal Barcode Printer GX430t, Model no. YSW 1000 or YSW 2000, may have been sold as an optional accessory for Sensititre Aris HIQ system YV4000-VZ
Power supply unit for Zebra Printers used with the device can potentially overheat or cause a fire hazard.
- 2025-03-26FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline procedure kits labeled as: 1) EYE PACK, Pack Number VAL038BSCLC
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
- 2025-03-26FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMedline procedure kits labeled as: 1) C SECTION, Pack Number DYNJ909098
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
- 2025-03-26FDA-DeviceBrainlab AGClass IIOrigin Data Management software versions 3.1.0, 3.1.1, 3.1.2, 3.2.0, 3.2.1
Under certain circumstances, it can occur that patient records from unrelated individuals are automatically merged into one patient record without user notification.
- 2025-03-26FDA-DeviceAngiodynamics, Inc.Class IIIsoLoc Short Profile Prostate Immobilization Treatment Balloon Device Kit; UPN/Catalog Number: ISO-60;
Product is mislabeled with the incorrect fill volume.
- 2025-03-26FDA-DeviceImmuno-Mycologics, IncClass IICandida ID Antigen, REAG, AG, REF C50110, Vol 1 mL REF ID1001, Immy Fungal Antigens, Positive Controls, & Immunodiffusion (ID) plates for use in the ID test, RX only, IVD, CE
Candida Immunodiffusion Antigen may have reduced sensitivity which could result in false negative results.
- 2025-03-26FDA-DeviceBeckman Coulter, Inc.Class IIDxI 9000 Access Immunoassay Analyzer, Catalog Number C11137.
the DxI 9000 encoder wheel was misaligned on the motor shaft, leading to increased slippage errors and causing the instrument to stop, posing a high safety risk due to potential delays in patient results.
- 2025-03-26FDA-DeviceBeckman Coulter, Inc.Class IIDxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278
Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
- 2025-03-26FDA-DeviceBeckman Coulter, Inc.Class IIAccess 2 Immunoassay Analyzer, Catalog Number 81600N
Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Infant, PT00156250 - Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 2m/6.5ft Extended Duration, MVIIH; Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 2m/6.5ft High Humidity, MVIIHH; Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 4m/13ft Extended Duration, MVIIHL; Microstream Advance Neonatal-Infant Intubated CO2 Filter
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Adult-Pediatric, PT00156254 - Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, MVAI; Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, MVAI100U; Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Extended Duration, MVAIH; Micr
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Luer Intubated Adult Pediatric, PT00156355 - Microstream Luer Adult-Pediatric Intubated CO2 Sampling Line 2m/6.5ft Short-term use: Procedural/Emergency, MLAI
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Infant Intubated, PT00134354 - Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 2m/6.5ft Extended Duration, QMVIIH
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Adult-Pediatric Intubated, PT00134333 - Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, QMVAI; Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Extended Duration, QMVAIH
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: OmniVentLine Set, 012808 - OmniVentLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter, 012495;
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Airway Adapter and CO2 Sampling Line, 01158 - Microstream Luer Adult-Pediatric Airway Adapter, 010989; Omnistream CO2 Sampling Line X25, 010991
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal, 10129497 - CapnoLine Neonatal-Infant Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), 006324COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology Length 7 (2.0 m), XS04620COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), XS04624COV
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
- 2025-03-26FDA-DeviceOridion Medical 1987 Ltd.Class IIMicrostream Instructions for Use and Part Number used with - Product Description, REF: FilterLine Sets & VitaLine Sets Adult/Pediatric, MQ04616 - MICROSTREAM FilterLine H Set Long Adult-Pediatric CO2 Sampling Line and Airway Adapter for Humid Environments, 007737; MICROSTREAM FilterLine Set Long Adult-Pediatric CO2 Sampling Line and Airway Adapter, 007768; MICROSTREAM FilterLine Set Adult-Pediatric CO2 Sampling Line and Airway Adapter, 010579; MICROSTREAM FilterLine H Set Adult-
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.