Moderate riskFDA DeviceClass II

Roche cobas c 303 analyzer recalled for vacuum tubing failure risk

Roche Diagnostics is recalling cobas c 303 analytical units because vacuum tubing in the cell rinse unit may deteriorate early from excess bending stress. The tubing may develop pinholes, cracks or ruptures, and fluid may drop into reaction cells without any system alarm, which could lead to erroneous patient and quality control results.

From the FDA notice · Read the official notice ↗ · All Roche Diagnostics recalls

What to do

  1. Set aside affected units rather than running patient or quality control samples on them.
  2. Check whether your unit is covered: Material Number 09793194001, UDI/DI 07613336220153, with serial numbers up to and including 7696-10.
  3. The notice doesn't say how to get a remedy; contact Roche Diagnostics Operations, Inc. for next steps.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
1002 units
Illnesses reported
None reported
Manufactured in
9115 Hague Rd, Indianapolis, Indiana, United States
Classification
Class II
Recall number
Z-0481-2027

In the FDA's words

Description
Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer Premature deterioration of the vacuum tubing in the cell rinse unit due to excess bending stress, which may cause pinholes, cracks or tubing rupture, and fluid to drop into random reaction cells without any system alarm, increasing the risk of erroneous patient and QC results. UDI/DI 07613336220153, All units up to and including SN (Serial Number) 7696-10 Distribution: US Nationwide distribution and territory of PR. Affected quantity: 1002 units
Hazard
Premature deterioration of the vacuum tubing in the cell rinse unit due to excess bending stress, which may cause pinholes, cracks or tubing rupture, and fluid to drop into random reaction cells without any system alarm, increasing the risk of erroneous patient and QC results.

Related topics