Roche cobas c 303 analyzer recalled for vacuum tubing failure risk
Roche Diagnostics is recalling cobas c 303 analytical units because vacuum tubing in the cell rinse unit may deteriorate early from excess bending stress. The tubing may develop pinholes, cracks or ruptures, and fluid may drop into reaction cells without any system alarm, which could lead to erroneous patient and quality control results.
From the FDA notice · Read the official notice ↗ · All Roche Diagnostics recalls
What to do
- Set aside affected units rather than running patient or quality control samples on them.
- Check whether your unit is covered: Material Number 09793194001, UDI/DI 07613336220153, with serial numbers up to and including 7696-10.
- The notice doesn't say how to get a remedy; contact Roche Diagnostics Operations, Inc. for next steps.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 1002 units
- Illnesses reported
- None reported
- Manufactured in
- 9115 Hague Rd, Indianapolis, Indiana, United States
- Classification
- Class II
- Recall number
- Z-0481-2027
In the FDA's words
- Description
- Roche cobas c 303 analytical unit, Material Number 09793194001, clinical chemistry analyzer Premature deterioration of the vacuum tubing in the cell rinse unit due to excess bending stress, which may cause pinholes, cracks or tubing rupture, and fluid to drop into random reaction cells without any system alarm, increasing the risk of erroneous patient and QC results. UDI/DI 07613336220153, All units up to and including SN (Serial Number) 7696-10 Distribution: US Nationwide distribution and territory of PR. Affected quantity: 1002 units
- Hazard
- Premature deterioration of the vacuum tubing in the cell rinse unit due to excess bending stress, which may cause pinholes, cracks or tubing rupture, and fluid to drop into random reaction cells without any system alarm, increasing the risk of erroneous patient and QC results.