Moderate riskFDA DeviceClass II

Hologic Selenia Dimensions systems recalled for potential freefall

Hologic is recalling Selenia Dimensions mammography systems because significant wear on the vertical lift nut of the C-arm assembly may cause sudden, unexpected freefall. The notice covers 16,089 units and reports no injuries.

From the FDA notice · Read the official notice ↗ · All Hologic recalls

What to do

  1. Notify your clinical or biomedical engineering staff so they can take the affected systems out of service until they have guidance from Hologic.
  2. Check the model numbers, catalog number P1979576, UDI-DI 9321502069847 and serial numbers listed in the notice against your systems.
  3. The notice does not say how to get a repair or replacement, so contact Hologic directly for instructions.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Selenia Dimensions Mammography Systems; System Name (Model/Catalog Number): 1. Selenia Dimensions 2D System Avia 3000/ (SDA-SYS-3000-2D-HTC), 2. Selenia Dimensions 2D System Avia 3000/ (SDA-SYS-3000-3D), 3. Selenia Dimensions 2D System 6000/ (SDM-SYS-6000-2D), 4. Selenia Dimensions 3D System 6000/ (SDM-SYS-6000-3D), 5. Selenia Dimensions Mammography System, Full Field Digital Mammography System, 2D (SDM-00001-2D), 6. Selenia Dimensions Mammography System 3D (SDM-00001-3D), 7. Selenia Dimensions Mammography System (SDM-00001-M2D), 8. Selenia Dimensions Mammography System (SDM-00001-M3D), 9. Selenia Dimensions FFDM System, 2D Avia, Fixed Height 5000, w/2MP Grayscale (SDM-05000-2A2), 10. Selenia Dimensions FFDM System, 2D Avia, Fixed Height 5000, w/3MP Grayscale (SDM-05000-2A3), 11. Selenia Dimensions FFDM System, 2D Avia, Fixed Height 5000, w/2MP Color (SDM-05000-2AC), 12. Selenia Dimensions Mammography System, Standard, 2D w/2MP Grayscale (SDM-05000-2D2), 13. Selenia Dimensions Mammography System, Standard, 2D w/3MP Grayscale (SDM-05000-2D3), 14. Selenia Dimensions Mammography System, Standard, 2D w/2MP Color (SDM-05000-2DC), 15. Selenia Dimensions Mammography System, Standard, 3D w/2MP Grayscale (SDM-05000-3D2), 16. Selenia Dimensions Mammography System, Standard, 3D w/3MP Grayscale (SDM-05000-3D3), 17. Selenia Dimensions Mammography System, Standard, 3D w/2MP Color (SDM-05000-3DC), 18. Selenia Dimensions 9000 System 2D (SDM-SYS-9000-2D), 19. Selenia Dimensions 9000 System 3D (SDM-SYS-9000-3D), 20. Selenia Dimensions Mammography System, 2D (SDA-SYS-3000-2D), 21. Selenia Dimensions 2D 9000 Package (SDM-SYS-9000-3D-HTC), 22. Selenia Dimensions 2D System 6000 (SDM-SYS-6000-2D-HTC), 23. Selenia Dimensions 3D System 6000 (SDM-SYS-6000-3D-HTC); Software Version: N/A Product Description: Selenia Dimensions Mammography System Component: N/A
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 2 products — see all of them.

Details

Units affected
16089
Illnesses reported
None reported
Manufactured in
600 Technology Dr, Newark, Delaware, United States
Classification
Class II
Recall number
Z-0520-2027

In the FDA's words

Description
Selenia Dimensions Mammography Systems; System Name (Model/Catalog Number): 1. Selenia Dimensions 2D System Avia 3000/ (SDA-SYS-3000-2D-HTC), 2. Selenia Dimensions 2D System Avia 3000/ (SDA-SYS-3000-3D), 3. Selenia Dimensions 2D System 6000/ (SDM-SYS-6000-2D), 4. Selenia Dimensions 3D System 6000/ (SDM-SYS-6000-3D), 5. Selenia Dimensions Mammography System, Full Field Digital Mammography System, 2D (SDM-00001-2D), 6. Selenia Dimensions Mammography System 3D (SDM-00001-3D), 7. Selenia Dimensions Mammography System (SDM-00001-M2D), 8. Selenia Dimensions Mammography System (SDM-00001-M3D), 9. Selenia Dimensions FFDM System, 2D Avia, Fixed Height 5000, w/2MP Grayscale (SDM-05000-2A2), 10. Selenia Dimensions FFDM System, 2D Avia, Fixed Height 5000, w/3MP Grayscale (SDM-05000-2A3), 11. Selenia Dimensions FFDM System, 2D Avia, Fixed Height 5000, w/2MP Color (SDM-05000-2AC), 12. Selenia Dimensions Mammography System, Standard, 2D w/2MP Grayscale (SDM-05000-2D2), 13. Selenia Dimensions Mammography System, Standard, 2D w/3MP Grayscale (SDM-05000-2D3), 14. Selenia Dimensions Mammography System, Standard, 2D w/2MP Color (SDM-05000-2DC), 15. Selenia Dimensions Mammography System, Standard, 3D w/2MP Grayscale (SDM-05000-3D2), 16. Selenia Dimensions Mammography System, Standard, 3D w/3MP Grayscale (SDM-05000-3D3), 17. Selenia Dimensions Mammography System, Standard, 3D w/2MP Color (SDM-05000-3DC), 18. Selenia Dimensions 9000 System 2D (SDM-SYS-9000-2D), 19. Selenia Dimensions 9000 System 3D (SDM-SYS-9000-3D), 20. Selenia Dimensions Mammography System, 2D (SDA-SYS-3000-2D), 21. Selenia Dimensions 2D 9000 Package (SDM-SYS-9000-3D-HTC), 22. Selenia Dimensions 2D System 6000 (SDM-SYS-6000-2D-HTC), 23. Selenia Dimensions 3D System 6000 (SDM-SYS-6000-3D-HTC); Software Version: N/A Product Description: Selenia Dimensions Mammography System Component: N/A The potential risk of sudden, unexpected freefall due to significant wear on the vertical lift nut of the C-Arm assembly on affected systems. Lot# E243A001; Exp. 02/28/2027 Distribution: Worldwide - US Nationwide TBD. Affected quantity: 16089
Hazard
The potential risk of sudden, unexpected freefall due to significant wear on the vertical lift nut of the C-Arm assembly on affected systems.

2 products in this recall

Hologic Selenia and 3Dimensions mammography systems recalled for freefall risk

Hologic is recalling two mammography system lines, Selenia Dimensions and 3Dimensions System 2D International, for a potential risk of sudden, unexpected freefall of the C-Arm. The risk comes from significant wear on the vertical lift nut of the C-Arm assembly. The notice lists 16,089 affected Selenia Dimensions units and 7,115 affected 3Dimensions units, and reports no injuries.

The agency lists each product separately. Check every one you might have.