Moderate riskFDA DeviceClass II

Hologic 3Dimensions mammography systems recalled for freefall risk

Hologic is recalling certain 3Dimensions mammography systems because of a potential risk of sudden, unexpected freefall. The risk comes from significant wear on the vertical lift nut of the C-Arm assembly. The FDA has classified this as a Class II recall, and no injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Hologic recalls

What to do

  1. Contact Hologic before continuing to use affected systems; the notice gives no interim instructions.
  2. Check whether your system is one of the nine model numbers in the notice (3DM-SYS-INTL2D through 3DM-SYS-STD-NS), and check for lots 222-08183870 (exp. 12/08/2026) or 222-08672499 (exp. 04/06/2026).
  3. The notice doesn't say how to get a repair or replacement; contact Hologic Inc.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Generic Name: 3DIMENSIONS MAMMOGRAPHY SYSTEM; System Name (Model/Catalog Number): 1. 3Dimensions System 2D International//(3DM-SYS-INTL2D), 2. 3Dimensions System 2D International Mobile//(3DM-SYS-INTL2D-MOB), 3. 3Dimensions System 2D International No Shield//(3DM-SYS-INTL2D-NS), 4. 3Dimensions System 3D International//(3DM-SYS-INTL3D), 5. 3Dimensions System 3D International Mobile//(3DM-SYS-INTL3D-MOB), 6. 3Dimensions System 3D International No Shield/(3DM-SYS-INTL3D-NS), 7. 3Dimensions System 3D Standard/(3DM-SYS-STD), 8. 3Dimensions System 3D Standard Mobile/ (3DM-SYS-STD-MOB), 9. 3Dimensions System 3D Standard No Shield/(3DM-SYS-STD-NS); Software Version: N/A Product Description: 3Dimensions Mammography System Component: N/A
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 2 products — see all of them.

Details

Units affected
7115
Illnesses reported
None reported
Manufactured in
600 Technology Dr, Newark, Delaware, United States
Classification
Class II
Recall number
Z-0521-2027

In the FDA's words

Description
Generic Name: 3DIMENSIONS MAMMOGRAPHY SYSTEM; System Name (Model/Catalog Number): 1. 3Dimensions System 2D International//(3DM-SYS-INTL2D), 2. 3Dimensions System 2D International Mobile//(3DM-SYS-INTL2D-MOB), 3. 3Dimensions System 2D International No Shield//(3DM-SYS-INTL2D-NS), 4. 3Dimensions System 3D International//(3DM-SYS-INTL3D), 5. 3Dimensions System 3D International Mobile//(3DM-SYS-INTL3D-MOB), 6. 3Dimensions System 3D International No Shield/(3DM-SYS-INTL3D-NS), 7. 3Dimensions System 3D Standard/(3DM-SYS-STD), 8. 3Dimensions System 3D Standard Mobile/ (3DM-SYS-STD-MOB), 9. 3Dimensions System 3D Standard No Shield/(3DM-SYS-STD-NS); Software Version: N/A Product Description: 3Dimensions Mammography System Component: N/A The potential risk of sudden, unexpected freefall due to significant wear on the vertical lift nut of the C-Arm assembly on affected systems. Various lots (see firm press release: https://www.prnewswire.com/news-releases/wedgewood-pharmacy-issues-recall-of-certain-compounded-veterinary-medications-within-expiry-due-to-potential-potency-issues-302915780.html); Exp Date Range: 10/18/2026 - 03/17/2027 Distribution: Worldwide - US Nationwide TBD. Affected quantity: 7115
Hazard
The potential risk of sudden, unexpected freefall due to significant wear on the vertical lift nut of the C-Arm assembly on affected systems.

2 products in this recall

Hologic Selenia and 3Dimensions mammography systems recalled for freefall risk

Hologic is recalling two mammography system lines, Selenia Dimensions and 3Dimensions System 2D International, for a potential risk of sudden, unexpected freefall of the C-Arm. The risk comes from significant wear on the vertical lift nut of the C-Arm assembly. The notice lists 16,089 affected Selenia Dimensions units and 7,115 affected 3Dimensions units, and reports no injuries.

The agency lists each product separately. Check every one you might have.