FDA-Device2022-12-28Class II
B105M Patient Monitor, REF 6160000-003
Hazard
GE Healthcare has become aware that if certain B1x5P / B1x5M patient monitors use "mmHg" or "kPa" as the unit for CO2 measurement, the displayed value for CO2 measurement can be inaccurate when used in a location that is not at or near sea level.
Sold states
Worldwide distribution.
Affected count
3905 units
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
B105M Patient Monitor, REF 6160000-003
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0731-2023More GE Medical Systems, LLC recalls
- FDA-Device2026-06-24GE Giraffe OmniBed infant bed recalled for detachable fastener hazard
- FDA-Device2026-06-24GE Giraffe OmniBed infant incubator recalled for loose fastener risk
- FDA-Device2026-06-24GE Giraffe OmniBed Carestation recalled for loose fastener risk in infant incubator
- FDA-Device2026-06-17GE HealthCare Allia IGS 3 Pulse angiographic X-ray system recalled for electrical shock risk to service personnel
- FDA-Device2026-06-17GE Allia IGS Pulse X-ray system recalled for electrical shock risk to service personnel