Moderate riskFDA DeviceClass II

GE Giraffe OmniBed infant bed recalled for detachable fastener hazard

GE HealthCare identified a potential issue with fasteners that secure the canopy soffit on certain Giraffe OmniBed infant beds. During transport or impact, one or more fasteners can come loose and fall into the patient compartment. In rare cases, multiple fasteners disengaging could cause the canopy soffit to loosen or detach entirely.

From the FDA notice · Read the official notice ↗ · All GE recalls

What to do

  1. Check if your device is one of the affected serial numbers listed in the recall notice.
  2. Contact GE HealthCare or your equipment supplier for instructions on fastener inspection or replacement.
  3. Inspect the canopy soffit and fasteners before use, especially after transport.
  4. Report any loose fasteners or canopy movement to GE HealthCare immediately.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 3 products — see all of them.

Details

Units affected
265 units
Illnesses reported
None reported
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Classification
Class II
Recall number
Z-2477-2026

In the FDA's words

Description
Giraffe OmniBed Field Replaceable Unit (FRU), Model Numbers 6600-1056-400 and 6600-1461-500 GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury. Serial Numbers: 5452896 801057486 800977619 801093759 5584569 5587625 5367527 5603110 5371240 5497382 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 265 units
Hazard
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

3 products in this recall

GE Giraffe OmniBed infant incubators recalled for loose fasteners

GE HealthCare is recalling three Giraffe OmniBed product lines, about 43,300 units in total, used as infant incubators and warmers. Under certain conditions, such as impact during transport, fasteners holding the canopy soffit beneath the control panel can come loose, and a fastener could fall into the patient compartment. In rare cases, the canopy soffit could become loose or detach, which could potentially injure a patient. No injuries have been reported.

The agency lists each product separately. Check every one you might have.