FDA-Device2026-06-24Class IIPROCESSING DEFECT

GE Giraffe OmniBed Carestation recalled for loose fastener risk in infant incubator

OTHERNationwide distribution

GE Giraffe OmniBed Carestation recalled for fastener hazard

GE HealthCare identified a potential issue with fasteners that secure the canopy soffit on certain Giraffe OmniBed Carestations. During transport or impact, one or more fasteners can come loose and fall into the patient compartment. In rare cases, multiple fasteners may disengage, causing the canopy soffit to loosen or detach.

  • Check your device serial number against the FDA notice to confirm if your unit is affected
  • Contact GE HealthCare or your equipment vendor for inspection and corrective action
  • Do not transport or move affected devices until fasteners have been checked and secured
Hazard

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.

Sold states
Worldwide - US Nationwide distribution.
Affected count
23430 units
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2476-2026