FDA-Device2026-01-14Class II

Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3

Securitas Healthcare LLC
Hazard

Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.

Sold states
Worldwide distribution -US Nationwide and the country of Canada.
Affected count
1898 units
Manufactured in
4600 Vine St, Lincoln, NE, United States
Products
Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1032-2026