Moderate riskFDA DeviceClass II

GE HealthCare SMV DSI nuclear medicine system recalled for detector fall hazard

GE HealthCare discovered that certain older SMV DSI nuclear medicine systems may have been moved or transported without proper detector support. This can stress the detector mounting and potentially cause the detector to fall, which could injure someone. No falls or injuries have occurred yet.

From the FDA notice · Read the official notice ↗ · All GE Healthcare recalls

What to do

  1. If your facility owns an SMV DSI system past its End of Guaranteed Service, contact GE HealthCare immediately to verify if your unit is affected
  2. Do not transport or relocate the system without first confirming proper detector support with GE HealthCare
  3. Inspect detector mounting mechanisms for any signs of damage or stress

Is yours affected?

Compare your label or packaging with the recalled product.

Product
SMV DSI, System, Tomography, Computed, Emission
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 4 products — see all of them.

Details

Units affected
3 units
Illnesses reported
None reported
Manufactured in
4, Hayozma St, Tirat Carmel, Israel
Classification
Class II
Recall number
Z-1045-2026

In the FDA's words

Description
SMV DSI, System, Tomography, Computed, Emission GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems. UDI/DI Not applicable: All serial numbers in distribution Distribution: US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece Affected quantity: 3 units
Hazard
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

4 products in this recall

GE HealthCare nuclear imaging systems recalled for detector fall risk

GE HealthCare is recalling four older nuclear medicine imaging systems, 60 units in total, that are past their end of guaranteed service. The systems may have been moved without proper detector support, which could stress the detector mounts and lead to a detector falling. No falls or injuries have been reported.

The agency lists each product separately. Check every one you might have.