Moderate riskFDA DeviceClass II

Medline pre-op kits recalled for stuck IV administration set check valves

Medline pre-op kits containing B. Braun IV administration sets have been recalled because the check valve components inside can get stuck in either the open or closed position. This could affect proper fluid delivery during medical procedures. No injuries have been reported to date.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check if you have any Medline pre-op kits with SKU numbers starting with DYKS (examples: DYKS1093B, DYKS1124C, DYKS1378B, or others listed in the recall)
  2. Do not use any affected kits; remove them from service immediately
  3. Contact Medline Industries or your supplier for instructions on returning the kits or obtaining replacements
  4. If kits were already used during a procedure, consult with your healthcare provider

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE-OP INPATIENT KIT, Medline kit SKU DYKS1379B; 13) PRE OP KIT, Medline kit SKU DYKS1382A; 14) PRE OP KIT, Medline kit SKU DYKS1416F; 15) PREOP KIT, Medline kit SKU DYKS1448B; 16) PRE-OP KIT, Medline kit SKU DYKS1458F; 17) PRE OP KIT, Medline kit SKU DYKS1467B; 18) PRE OP KIT, Medline kit SKU DYKS1468; 19) PRE-OP, Medline kit SKU DYKS1482A; 20) ADULT PRE OP, Medline kit SKU DYKS1487I; 21) PRE OP ENDO, Medline kit SKU DYKS1489F; 22) PREOP KIT, Medline kit SKU DYKS1507B; 23) PRE OP KIT, Medline kit SKU DYKS1518A; 24) PRE OP KIT, Medline kit SKU DYKS1544F; 25) PREOP, Medline kit SKU DYKS1551B; 26) PREOP, Medline kit SKU DYKS1552B; 27) PREOP, Medline kit SKU DYKS1580; 28) PRE OP KIT, Medline kit SKU DYKS2024A; 29) PRE OP KIT, Medline kit SKU DYKS2056A; 30) PRE-OP KIT, Medline kit SKU DYKS2074; 31) PRE-OP KIT, Medline kit SKU DYKS2074A; 32) PREOP KIT, Medline kit SKU DYKS2076; 33) PRE OP KIT 1, Medline kit SKU DYKS2083; 34) PRE OP KIT, Medline kit SKU DYKS2103.
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 12 products — see all of them.

Details

Units affected
11173 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1057-2026

In the FDA's words

Description
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) PRE OP KIT, Medline kit SKU DYKS1093B; 2) PRE OP KIT, Medline kit SKU DYKS1124C; 3) PRE OP KIT, Medline kit SKU DYKS1128C; 4) PREOP KIT, Medline kit SKU DYKS1130B; 5) PRE OP KIT, Medline kit SKU DYKS1163G; 6) PRE OP KIT SURGERY, Medline kit SKU DYKS1263G; 7) PREOP KIT, Medline kit SKU DYKS1313A; 8) PRE OP KIT, Medline kit SKU DYKS1325A; 9) ADULT PRE OP KIT, Medline kit SKU DYKS1337D; 10) PRE OP KIT ADULT, Medline kit SKU DYKS1362A; 11) PRE-OP - OUTPATIENT KIT, Medline kit SKU DYKS1378B; 12) PRE-OP INPATIENT KIT, Medline kit SKU DYKS1379B; 13) PRE OP KIT, Medline kit SKU DYKS1382A; 14) PRE OP KIT, Medline kit SKU DYKS1416F; 15) PREOP KIT, Medline kit SKU DYKS1448B; 16) PRE-OP KIT, Medline kit SKU DYKS1458F; 17) PRE OP KIT, Medline kit SKU DYKS1467B; 18) PRE OP KIT, Medline kit SKU DYKS1468; 19) PRE-OP, Medline kit SKU DYKS1482A; 20) ADULT PRE OP, Medline kit SKU DYKS1487I; 21) PRE OP ENDO, Medline kit SKU DYKS1489F; 22) PREOP KIT, Medline kit SKU DYKS1507B; 23) PRE OP KIT, Medline kit SKU DYKS1518A; 24) PRE OP KIT, Medline kit SKU DYKS1544F; 25) PREOP, Medline kit SKU DYKS1551B; 26) PREOP, Medline kit SKU DYKS1552B; 27) PREOP, Medline kit SKU DYKS1580; 28) PRE OP KIT, Medline kit SKU DYKS2024A; 29) PRE OP KIT, Medline kit SKU DYKS2056A; 30) PRE-OP KIT, Medline kit SKU DYKS2074; 31) PRE-OP KIT, Medline kit SKU DYKS2074A; 32) PREOP KIT, Medline kit SKU DYKS2076; 33) PRE OP KIT 1, Medline kit SKU DYKS2083; 34) PRE OP KIT, Medline kit SKU DYKS2103. Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions. Medline kit SKU DYKS1378B, UDI/DI 10195327521660 (EA) 40195327521661 (CS), Lot Number 25HBB410; Medline kit SKU DYKS1448B, UDI/DI 10195327691349 (EA) 40195327691340 (CS), Lot Number 25HBE055; Medline kit SKU DYKS1163G, UDI/DI 10195327503185 (EA) 40195327503186 (CS), Lot Number 25HBF057; Medline kit SKU DYKS1487I, UDI/DI 10198459415678 (EA) 40198459415679 (CS), Lot Number 25HBI512; Medline kit SKU DYKS2056A, UDI/DI 10198459467493 (EA) 40198459467494 (CS), Lot Number 25HBK906; Medline kit SKU DYKS2074, UDI/DI 10198459399893 (EA) 40198459399894 (CS), Lot Number 25HBV987; Medline kit SKU DYKS2083, UDI/DI 10198459435126 (EA) 40198459435127 (CS), Lot Number 25HLA076; Medline kit SKU DYKS1552B, UDI/DI 10198459269523 (EA) 40198459269524 (CS), Lot Number 25HLA432; Medline kit SKU DYKS1362A, UDI/DI 10195327343163 (EA) 40195327343164 (CS), Lot Number 25HLA829; Medline kit SKU DYKS1124C, UDI/DI 10198459447877 (EA) 40198459447878 (CS), Lot Number 25HMB303; Medline kit SKU DYKS1313A, UDI/DI 10198459370595 (EA) 40198459370596 (CS), Lot Number 25HMB948; Medline kit SKU DYKS1544F, UDI/DI 10198459411700 (EA) 40198459411701 (CS), Lot Number 25HMC488; Medline kit SKU DYKS1507B, UDI/DI 10198459489372 (EA) 40198459489373 (CS), Lot Number 25HMD440; Medline kit SKU DYKS1124C, UDI/DI 10198459447877 (EA) 40198459447878 (CS), Lot Number 25HMF480; Medline kit SKU DYKS1130B, UDI/DI 10198459377129 (EA) 40198459377120 (CS), Lot Number 25IBC508; Medline kit SKU DYKS1378B, UDI/DI 10195327521660 (EA) 40195327521661 (CS), Lot Number 25IBI706; Medline kit SKU DYKS1551B, UDI/DI 10198459347702 (EA) 40198459347703 (CS), Lot Number 25IBK720; Medline kit SKU DYKS1580, UDI/DI 10198459024320 (EA) 40198459024321 (CS), Lot Number 25IBL590; Medline kit SKU DYKS2076, UDI/DI 10198459399732 (EA) 40198459399733 (CS), Lot Number 25IBL885; Medline kit SKU DYKS1163G, UDI/DI 10195327503185 (EA) 40195327503186 (CS), Lot Number 25IBQ969; Medline kit SKU DYKS1518A, UDI/DI 10198459032448 (EA) 40198459032449 (CS), Lot Number 25IBU385; Medline kit SKU DYKS1489F, UDI/DI 10198459291272 (EA) 40198459291273 (CS), Lot Number 25IBW167; Medline kit SKU DYKS1337D, UDI/DI 10198459078415 (EA) 40198459078416 (CS), Lot Number 25ILA943; Medline kit SKU DYKS1468, UDI/DI 10195327411213 (EA) 40195327411214 (CS), Lot Number 25ILA944; Medline kit SKU DYKS1467B, UDI/DI 10198459025648 (EA) 40198459025649 (CS), Lot Number 25IME692; Medline kit SKU DYKS1379B, UDI/DI 10195327387433 (EA) 40195327387434 (CS), Lot Number 25JBB244; Medline kit SKU DYKS1551B, UDI/DI 10198459347702 (EA) 40198459347703 (CS), Lot Number 25JBC342; Medline kit SKU DYKS1551B, UDI/DI 10198459347702 (EA) 40198459347703 (CS), Lot Number 25JBE269; Medline kit SKU DYKS2076, UDI/DI 10198459399732 (EA) 40198459399733 (CS), Lot Number 25JBE865; Medline kit SKU DYKS1128C, UDI/DI 10195327549374 (EA) 40195327549375 (CS), Lot Number 25JBG502; Medline kit SKU DYKS1487I, UDI/DI 10198459415678 (EA) 40198459415679 (CS), Lot Number 25JBJ439; Medline kit SKU DYKS2076, UDI/DI 10198459399732 (EA) 40198459399733 (CS), Lot Number 25JBJ451; Medline kit SKU DYKS1263G, UDI/DI 101984
Hazard
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

12 products in this recall

Medline procedure kits recalled for IV check valve problem

Medline is recalling 12 kits that contain B. Braun IV administration sets and pump administration sets. Customers reported that the check valve components in these sets can get stuck in the open or closed position. No injuries have been reported.

The agency lists each product separately. Check every one you might have.