Beckman Coulter AU5800 chemistry analyzer recalled for software errors causing delayed test results
Contact Beckman Coulter about your AU5800 analyzer
Beckman Coulter AU5800 chemistry analyzers with certain software configurations may experience errors that prevent the system from opening calibration databases after running for more than 15 minutes. When this happens, the analyzer stops working and requires a service technician to fix it, which can delay patient test results and treatment.
- If you operate an AU5800 analyzer, contact Beckman Coulter service immediately to determine if your unit is affected.
- Do not ignore database error messages—they indicate the problem has occurred.
- Schedule a service visit to apply the software fix before the system fails during patient testing.
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1553-2026More Beckman Coulter Mishima K.K. recalls