CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Beckman Coulter Mishima K.K.
Hazard
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment
Sold states
US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.
Affected count
47
Manufactured in
454-32, Higashino; Nagaizumi-Cho, Sunto-Gun, N/A, Japan
Products
CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1553-2026

Original summary

CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281 A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment REF/UDI-DI/ Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B96697/14987666543016/2025061442, 2025061446, 2025061452, 2025061454, 2025061464, 2025071479, 2025071488, 2025081513, 2025081516, 2025091536, 2025091538, 2025091543, 2025091559, 2025091561, 2025111598, 2025111602, 2025111604, 2025111605, 2025111606, 2025121618, 2025121622, 2025121623. B96698/14987666543023/2025061449, 2025061451, 2025061457, 2025061467, 2025071475, 2025071476, 2025071478, 2025071481, 2025071495, 2025071497, 2025081503, 2025081524, 2025091539, 2025091549, 2025091551. Older Instrument Replaced with PC preinstalled with Microsoft SQL Server 2022 Express: B23279/14987666541296/2014011316, 2014071647, 2014121858, 2014121906. B23280/14987666541302/2016022593. B23281/14987666541319/5094354. B96697/14987666543016/2019074683. B96698/14987666543023/2017063377, 2019114891, 2024101288 Distribution: US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR. Affected quantity: 47