FDA-Device2025-07-30Class II
SIGNA Architect AIR
Hazard
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Sold states
Worldwide distribution - US Nationwide and the country of China.
Affected count
N/A
Manufactured in
3200 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
SIGNA Architect AIR
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2131-2025More GE Medical Systems, LLC recalls
- FDA-Device2026-07-22GE Optima XR200/220/240amx radiography systems recalled for potential arm drop hazard
- FDA-Device2026-06-24GE Giraffe OmniBed infant bed recalled for detachable fastener hazard
- FDA-Device2026-06-24GE Giraffe OmniBed infant incubator recalled for loose fastener risk
- FDA-Device2026-06-24GE Giraffe OmniBed Carestation recalled for loose fastener risk in infant incubator
- FDA-Device2026-06-17GE HealthCare Allia IGS 3 Pulse angiographic X-ray system recalled for electrical shock risk to service personnel