Medline reprocessed EP catheters recalled for residual particle risk
Specific lots of Medline reprocessed St. Jude Medical Response diagnostic electrophysiology catheters may contain small particles of residual material on surfaces that contact patients. If these particles dislodge during use, they could potentially cause blood clots, inflammation, or infection.
From the FDA notice · Read the official notice ↗ · All Medline recalls
What to do
- Check if your facility has catheters from lot EP250203 (item numbers 401206RH or 401305RH).
- Do not use affected catheters.
- Contact Medline Industries for instructions on returning or replacing the devices.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.
- Sold in
- Nationwide
- Made for
- All ages
Details
- Units affected
- 3
- Illnesses reported
- None reported
- Manufactured in
- 3 Lakes Dr, N/A, Northfield, IL, United States
- Classification
- Class I
- Recall number
- Z-2615-2026
In the FDA's words
- Description
- Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm. Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Lot EP250203 Item no. 401206RH - UDI-DI 10197344043767; Item no. 401305RH - UDI-DI 10197344043637. Distribution: US Nationwide distribution in the state of TX. Affected quantity: 3
- Hazard
- Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
More from Medline
See all- Medline Brand beds recalled for possible fire and entrapment risk
SeriousFDANov 25, 2026
- Medline convenience kits recalled for discolored IV drip chambers
ModerateFDANov 25, 2026
- Medline SENSICARE PI surgical gloves recalled, not sterilized
ModerateFDANov 25, 2026
- Medline surgical procedure kits recalled for reprocessed components
ModerateFDANov 20, 2026
- Medline Bubble CPAP kit recalled for possible heat damage risk
ModerateFDANov 20, 2026