Moderate riskFDA DeviceClass II

Encore Medical EMPOWR 3D knee inserts recalled for packaging errors

Encore Medical is recalling certain EMPOWR 3D KNEE INS inserts because a packaging discrepancy mixed up product sizes. The notice reports no injuries, and it lists 8 affected units sent to distributors and medical providers in several states.

From the FDA notice · Read the official notice ↗ · All Encore Medical recalls

What to do

  1. Set aside any EMPOWR 3D KNEE INS, 11L 14MM, VE (catalog 341-14-711) from lots 145T1067 or 135T1060, and check the size on each package before use.
  2. Check the GTIN 00888912167086 and expiration dates on the notice: 12-06-2026 for lot 145T1067 and 12-07-2026 for lot 135T1060.
  3. Contact Encore Medical for next steps, because the notice does not specify a remedy.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 341-14-711 Product Description: Enovis knee devices are for knee arthroplasty.
Sold in
GL112 Certus Medical LLC Indiana (IN) & Michigan (MI) 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 MW001 Fox Medtech, LLC Colorado (CO), Wyoming (WY), Montana (MT), Counties in Nebraska 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 NE190 Implant Medical Sales, Inc. New York (NY) 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 SE180 Joint Medical Solutions, L.L.C. Alabama (AL) 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 GL108 Great Lakes Medical Illiniois (IL) 341-14-711 135T1060 EMPOWR 3D KNEETM INS, 11L 14MM, VE 2 GL112 Certus Medical LLC Indiana (IN) & Michigan (MI) 341-14-711 135T1060 EMPOWR 3D KNEETM INS, 11L 14MM, VE 1 CE120 Matrix Orthopedics Surgical Devices Dallas/Fort Worth, TX 342-14-711 145T1067 EMPOWR 3D KNEETM INS, 11R 14MM, VE 1 NW105 Kinetic Medicine South Dakota (SD), Eastern Nebraska (NE), Western Iowa (IA) 342-14-711 145T1067 EMPOWR 3D KNEETM INS, 11R 14MM, VE 2 MA102 ProCore Medical LLC - NC North Carolina (NC) 342-14-711 145T1067 EMPOWR 3D KNEETM INS, 11R 14MM, VE 1 Total 11
Made for
Adults (18 years to 100 years)

This recall covers 2 products — see all of them.

Details

Units affected
8
Illnesses reported
None reported
Manufactured in
9800 Metric Blvd, N/A, Austin, TX, United States
Classification
Class II
Recall number
Z-2701-2026

In the FDA's words

Description
Brand Name: EMPOWR 3D KNEE Product Name: EMPOWR 3D KNEE INS, 11L 14MM, VE Model/Catalog Number: 341-14-711 Product Description: Enovis knee devices are for knee arthroplasty. Due to packaging discrepancy with incorrect product size mixed up. Model/Catalog Number: 341-14-711 Lot Code: Lot: 145T1067 and 135T1060 GTIN: 00888912167086 Expiration Date: 12-06-2026 (145T1067) 12-07-2026 (135T1060) Distribution: GL112 Certus Medical LLC Indiana (IN) & Michigan (MI) 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 MW001 Fox Medtech, LLC Colorado (CO), Wyoming (WY), Montana (MT), Counties in Nebraska 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 NE190 Implant Medical Sales, Inc. New York (NY) 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 SE180 Joint Medical Solutions, L.L.C. Alabama (AL) 342-19-703 177T1057 EMPOWR 3D KNEETM INS, 3R 19MM, VE 1 GL108 Great Lakes Medical Illiniois (IL) 341-14-711 135T1060 EMPOWR 3D KNEETM INS, 11L 14MM, VE 2 GL112 Certus Medical LLC Indiana (IN) & Michigan (MI) 341-14-711 135T1060 EMPOWR 3D KNEETM INS, 11L 14MM, VE 1 CE120 Matrix Orthopedics Surgical Devices Dallas/Fort Worth, TX 342-14-711 145T1067 EMPOWR 3D KNEETM INS, 11R 14MM, VE 1 NW105 Kinetic Medicine South Dakota (SD), Eastern Nebraska (NE), Western Iowa (IA) 342-14-711 145T1067 EMPOWR 3D KNEETM INS, 11R 14MM, VE 2 MA102 ProCore Medical LLC - NC North Carolina (NC) 342-14-711 145T1067 EMPOWR 3D KNEETM INS, 11R 14MM, VE 1 Total 11 Affected quantity: 8
Hazard
Due to packaging discrepancy with incorrect product size mixed up.

2 products in this recall

EMPOWR 3D KNEE inserts recalled for packaging size mix-up

Encore Medical is recalling EMPOWR 3D KNEE knee inserts, which are used in knee replacement surgery, because the packaging may show the wrong product size. The recall covers two product listings, under catalog numbers 342-19-703 and 341-14-711, and the notice also lists 342-14-711 in its distribution records. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.