Moderate riskFDA DeviceClass II

Medtronic EOPA arterial cannula recalled for potential pinhole leaks

Pinhole leaks may occur in the wire-wound body of the EOPA 22FR arterial cannula during cardiopulmonary bypass. Potential risks may include low blood volume, vessel laceration or perforation, low blood pressure, ischemia, and organ dysfunction. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Medtronic recalls

What to do

  1. Set aside unused EOPA 22FR cannulas (Model 77422) whose lot number or UDI-DI matches the notice, and contact Medtronic Perfusion Systems for instructions.
  2. Affected UDI-DIs are 20763000135632 and 00763000135638; the matching lot numbers are listed in the notice.
  3. The notice doesn't specify a remedy; ask Medtronic Perfusion Systems how to return or replace affected units.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 6 products — see all of them.

Details

Units affected
21,290 units
Illnesses reported
None reported
Manufactured in
7611 Northland Dr N, N/A, Brooklyn Park, MN, United States
Classification
Class II
Recall number
Z-2806-2026

In the FDA's words

Description
EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422 Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction. UDI-DI: 20763000135632, Lot Numbers: 0232446429 0232446444 0232486752 0232601515 0232601516 0232601522 0232601586 0232601591 0232601593 0232602180 0232798265 0232798353 0232849583 0232894981 0232895082 0232902694 0232902697 0232902698 0232902700 0232902803 0232902805 0232902846 0232985469 0233009922 0233108251 0233116991 0233127352 0233334923 0233334966 0233426694 0233426695 0233528727; UDI-DI: 00763000135638, Lot Numbers: 0232446444 0232798353 0232895082 0232902697 0232902803 0232902846 0232985469 0233127352 0233334923 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 21,290 units
Hazard
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

6 products in this recall

Medtronic EOPA arterial cannulas recalled for potential pinhole leaks

Medtronic is recalling six EOPA arterial cannula models, used as arterial return cannulas during cardiopulmonary bypass. Pinhole leaks may occur in the wire-wound body during bypass, which may lead to hypovolemia, vessel laceration or perforation, hypotension, ischemia, or organ dysfunction. The notice reports no injuries.