Medtronic EOPA arterial cannula recalled for possible pinhole leaks
Medtronic Perfusion Systems is recalling the EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, Model 77722, which is used as an arterial return cannula during cardiopulmonary bypass. Pinhole leaks may occur in the wire-wound body during bypass. Possible risks include low blood volume, vessel laceration or perforation, low blood pressure, reduced blood flow to tissues, and organ dysfunction. No injuries have been reported.
From the FDA notice · Read the official notice ↗ · All Medtronic recalls
What to do
- Set aside affected lots and do not use them for cardiopulmonary bypass.
- Check UDI-DI 20199150023312 for lots 0232657469, 0232670800, 0232744945, 0232798410, 0232862769, 0232985251, 0233116924, and 0233116931.
- Check UDI-DI 00199150023318 for lot 0232744945.
- The notice does not specify a remedy; contact Medtronic Perfusion Systems for return or replacement instructions.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 6 products — see all of them.
Details
- Units affected
- 3,110 units
- Illnesses reported
- None reported
- Manufactured in
- 7611 Northland Dr N, N/A, Brooklyn Park, MN, United States
- Classification
- Class II
- Recall number
- Z-2809-2026
In the FDA's words
- Description
- EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722 Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction. UDI-DI: 20199150023312, Lot Numbers: 0232657469 0232670800 0232744945 0232798410 0232862769 0232985251 0233116924 0233116931; UDI-DI: 00199150023318, Lot Numbers: 0232744945 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 3,110 units
- Hazard
- Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
6 products in this recall
Medtronic EOPA arterial cannulas recalled for potential pinhole leaks
Medtronic is recalling six EOPA arterial cannula models, used as arterial return cannulas during cardiopulmonary bypass. Pinhole leaks may occur in the wire-wound body during bypass, which may lead to hypovolemia, vessel laceration or perforation, hypotension, ischemia, or organ dysfunction. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77422Z-2806-2026 · 21,290 units
- EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522Z-2807-2026 · 11,720 units
- EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622Z-2808-2026 · 13,510 units
- EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722 (this page)Z-2809-2026 · 3,110 units
- EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222Z-2810-2026 · 3,621 units
- EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322Z-2811-2026 · 1,424 units