Tissium Coaptium Connect nerve device recalled for shelf-life issue
Tissium SA is relabeling Coaptium Connect nerve repair devices (model NCOCO1575) because the labeled 36-month shelf life is not supported by real-time data. The devices will be relabeled with a 26-month shelf life. Using a device beyond its validated shelf life may cause it to break during surgery, which could require revision surgery.
From the FDA notice · Read the official notice ↗
What to do
- Set aside affected lots and do not use them beyond the revised 26-month shelf life.
- Identify affected devices by model NCOCO1575, UDI-DI 03760279380087, and lot numbers 250700007, 251000031, or 260300041.
- The notice does not say how to get relabeled devices or replacements; contact Tissium SA.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 7 products — see all of them.
Details
- Illnesses reported
- None reported
- Manufactured in
- 74 Rue Du Faubourg Saint-Antoine, Paris 12e Arrondissement, France
- Classification
- Class II
- Recall number
- Z-2902-2026
In the FDA's words
- Description
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. Lot Code: Model No NCOCO1575; UDI-DI ¿03760279380087; Lot Number 250700007, 251000031, and 260300041 Distribution: US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
- Hazard
- Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
7 products in this recall
The agency lists each product separately. Check every one you might have.
- Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizi…Z-2901-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizi… (this page)Z-2902-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizi…Z-2903-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizi…Z-2904-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizi…Z-2906-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizi…Z-2907-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizi…Z-2908-2026