Moderate riskFDA DeviceClass II

Tissium Coaptium Connect nerve device recalled for shelf-life issue

Tissium SA is relabeling Coaptium Connect nerve repair devices (model NCOCO1575) because the labeled 36-month shelf life is not supported by real-time data. The devices will be relabeled with a 26-month shelf life. Using a device beyond its validated shelf life may cause it to break during surgery, which could require revision surgery.

From the FDA notice · Read the official notice ↗

What to do

  1. Set aside affected lots and do not use them beyond the revised 26-month shelf life.
  2. Identify affected devices by model NCOCO1575, UDI-DI 03760279380087, and lot numbers 250700007, 251000031, or 260300041.
  3. The notice does not say how to get relabeled devices or replacements; contact Tissium SA.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 7 products — see all of them.

Details

Illnesses reported
None reported
Manufactured in
74 Rue Du Faubourg Saint-Antoine, Paris 12e Arrondissement, France
Classification
Class II
Recall number
Z-2902-2026

In the FDA's words

Description
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. Lot Code: Model No NCOCO1575; UDI-DI ¿03760279380087; Lot Number 250700007, 251000031, and 260300041 Distribution: US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Hazard
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.