Tissium Coaptium Connect recalled for unverified shelf life
Tissium SA is recalling Coaptium Connect nerve repair devices (model NCOCO2010) because the labeled 36-month shelf life is not supported by real-time data. The devices will be relabeled with a 26-month shelf life. Using a device past its validated shelf life may cause it to break during surgery, which could require revision surgery.
From the FDA notice · Read the official notice ↗
What to do
- Clinicians: check whether you have lots 250700042, 251000033 or 251000034 (model NCOCO2010, UDI-DI 03760279380094) and set aside any unused devices from these lots.
- Contact Tissium SA before using any affected device; the notice does not describe a replacement or refund.
- Patients with this device already implanted: contact your surgeon or care team. Do not take action on your own.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 7 products — see all of them.
Details
- Illnesses reported
- None reported
- Manufactured in
- 74 Rue Du Faubourg Saint-Antoine, Paris 12e Arrondissement, France
- Classification
- Class II
- Recall number
- Z-2903-2026
In the FDA's words
- Description
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery. Lot Code: Model No NCOCO2010; UDI-DI 03760279380094; Lot Number 250700042, 251000033, and 251000034 Distribution: US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
- Hazard
- Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
7 products in this recall
The agency lists each product separately. Check every one you might have.
- Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizi…Z-2901-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizi…Z-2902-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizi… (this page)Z-2903-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizi…Z-2904-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizi…Z-2906-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizi…Z-2907-2026
- Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizi…Z-2908-2026