Moderate riskFDA DeviceClass II

Tobii Dynavox TD Navio Midi speech device recalled for battery risk

The battery mounting tabs in some TD Navio Midi speech-generating devices may become damaged, allowing the battery to shift inside the device. This can cause an internal short circuit, leading to localized heating and gas venting inside the device enclosure.

From the FDA notice · Read the official notice ↗

What to do

  1. Check if your device serial number falls between TDNM1-240628000004 and TDNM1-260617070001.
  2. If affected, stop using the device and contact Tobii Dynavox for a replacement or repair.
  3. Do not attempt to repair or adjust the battery mounting yourself.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 2 products — see all of them.

Details

Illnesses reported
None reported
Manufactured in
2100 Wharton St Ste Ste 400, Pittsburgh, PA, United States
Classification
Class II
Recall number
Z-2959-2026

In the FDA's words

Description
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798 Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure. UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design). Distribution: US Nationwide distribution.
Hazard
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

2 products in this recall

Tobii Dynavox TD Navio devices recalled for battery risk

Tobii Dynavox is recalling two TD Navio speech generating devices: the TD Navio Midi (model 1000798) and the TD Navio Maxi (model 1000799). Under certain conditions, the battery mounting tabs can become damaged, letting the battery move inside the device and possibly causing an internal short circuit, localized heating, and gas venting. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

Related topics