GE Portrait Core Services Software V1.1 recalled for patient monitoring loss and alarm reset risk
What this means
GE Medical Systems Portrait Core Services Software V1.1 contains a software issue that may cause loss of patient monitoring at the Hub and Central Viewer. The update may also reset all alarm settings to factory defaults, overriding previously configured alarm settings for individual patients or care areas.
- Contact GE Medical Systems Information Technologies Inc for guidance on the software update
- Do not install the update until you receive corrective instructions
- Verify patient monitoring and alarm settings after any software changes
The agency's original wording
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3036-2026More GE recalls
- FDA-Device2026-09-23GE Carescape Central Station CSCS V3 recalled for patient monitoring delay after restart
- FDA-Device2026-09-02GE PORTRAIT Core Services software update may disable patient monitoring and reset alarms
- FDA-Device2026-09-02GE Portrait Compute and Core Services V1.2 software recalled for patient monitoring loss risk
- FDA-Device2026-09-02GE Portrait Core Services Software v1.0.5 recalled for patient monitoring loss and alarm reset risk