FDA-Device2026-09-23Class IIPROCESSING DEFECT

GE Carescape Central Station CSCS V3 recalled for patient monitoring delay after restart

OTHERNationwide distributionMedical devices

What this means

GE's Carescape Central Station patient monitoring system may take up to 3 minutes to resume monitoring after a restart. During this gap, clinically important changes in a patient's condition might not be detected right away.

  • If you operate this device, contact GE Medical Systems for guidance on the issue and any available updates.
  • Be aware of the potential 3-minute monitoring gap after system restarts and plan accordingly in your workflow.
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

Intended for
All ages
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution.
Affected count
29,570 (23,632 US; 5,938 OUS)
Manufactured in
3114 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3143-2026