GE Carescape Central Station CSCS V3 recalled for patient monitoring delay after restart
GE's Carescape Central Station patient monitoring system may take up to 3 minutes to resume monitoring after a restart. During this gap, clinically important changes in a patient's condition might not be detected right away.
From the FDA notice · Read the official notice ↗ · All GE recalls
What to do
- If you operate this device, contact GE Medical Systems for guidance on the issue and any available updates.
- Be aware of the potential 3-minute monitoring gap after system restarts and plan accordingly in your workflow.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
Details
- Units affected
- 29,570 (23,632 US; 5,938 OUS)
- Illnesses reported
- None reported
- Manufactured in
- 3114 N Grandview Blvd, N/A, Waukesha, WI, United States
- Classification
- Class II
- Recall number
- Z-3143-2026
In the FDA's words
- Description
- Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events. 5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV Distribution: Worldwide - US Nationwide distribution. Affected quantity: 29,570 (23,632 US; 5,938 OUS)
- Hazard
- Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.