ICU Medical oncology IV tubing sets recalled for black specks in drip chambers
ICU Medical oncology IV tubing sets may contain small black specks in the drip chamber. These specks are discolored plastic material from the manufacturing process and do not enter the fluid path. Healthcare providers who notice the specks may replace the set before use, which could delay treatment.
From the FDA notice · Read the official notice ↗ · All ICU Medical recalls
What to do
- If you use or dispense these sets, inspect drip chambers for black specks before patient use.
- Contact ICU Medical or your distributor for information on affected lot numbers and replacements.
- Do not delay patient therapy if a set is suspected; use an unaffected replacement immediately.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Oncology sets with drip chambers
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 8 products — see all of them.
Details
- Units affected
- 107,800
- Illnesses reported
- None reported
- Manufactured in
- 951 Calle Amanecer, N/A, San Clemente, CA, United States
- Classification
- Class II
- Recall number
- Z-3158-2026
In the FDA's words
- Description
- Oncology sets with drip chambers Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy. All products manufactured on or before 7/4/2026. UDI-DI: 10840619054486, 10887709091914 Distribution: Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR. Affected quantity: 107,800
- Hazard
- Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
8 products in this recall
ICU Medical IV sets recalled for particulate contamination risk
ICU Medical is recalling eight IV administration sets with drip chambers. The drip chambers may have black specks, which may come from discolored plastic left behind during manufacturing. The company has not ruled out discoloration or particles in the fluid path. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Plum administration setsZ-3156-2026 · 163,500,199
- IV Non-Dedicated Gravity Administration Sets with drip chambersZ-3157-2026 · 2,450
- Oncology sets with drip chambers (this page)Z-3158-2026 · 107,800
- Oncology sets with drip chambersZ-3159-2026 · 28,550
- IV Sets with drip chambersZ-3160-2026 · 100,712,657
- IV Sets with drip chambersZ-3161-2026 · 6,105,468
- IV Sets with drip chambersZ-3162-2026 · 86,113
- IV Sets with drip chambersZ-3163-2026 · 56,553,672