Moderate riskFDA DeviceClass II

ICU Medical oncology IV tubing sets recalled for black specks in drip chambers

ICU Medical oncology IV tubing sets may contain small black specks in the drip chamber. These specks are discolored plastic from the manufacturing process and do not enter the fluid path. Healthcare providers who notice them may replace the tubing set before use, which could delay patient treatment.

From the FDA notice · Read the official notice ↗ · All ICU Medical recalls

What to do

  1. Check oncology IV tubing sets for black specks in the drip chamber before use.
  2. If you see black specks, do not discard the set automatically — contact ICU Medical or your healthcare facility for guidance.
  3. Contact ICU Medical or your hospital's medical device safety team if you have questions about affected products.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Oncology sets with drip chambers
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 8 products — see all of them.

Details

Units affected
28,550
Illnesses reported
None reported
Manufactured in
951 Calle Amanecer, N/A, San Clemente, CA, United States
Classification
Class II
Recall number
Z-3159-2026

In the FDA's words

Description
Oncology sets with drip chambers Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy. All products manufactured on or before 7/4/2026. UDI-DI: 10887709076768, 10887709088105 Distribution: Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR. Affected quantity: 28,550
Hazard
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.

8 products in this recall