Howmedica Osteonics knee inserts recalled for possible excessive wear
Howmedica Osteonics manufactured multiple knee implant components (inserts, bearings, and patella components) using polyethylene material that was stored for more than five years before use. The aged material may have developed excessive oxidation, which can cause the implant to wear faster than intended and potentially break apart inside the body.
From the FDA notice · Read the official notice ↗ · All Howmedica Osteonics recalls
What to do
- If you have one of the affected implant components, contact your surgeon or healthcare provider immediately with your implant model number and lot number.
- Do not remove the implant yourself; work with your medical team to determine if replacement is needed.
- Keep implant documentation and lot numbers available for your healthcare provider.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PATELLA 73-2710 SCORPIO INSET PATELLA 73-2610 SCORPIO M-DOME PATELLA 73-0110 73-0710 73-0910 SCORPIO RECESSED PATELLA 26MM 3044-0026 SCORPIO TS TIB INS 72-4-9114 SCORPIO TS TIBIAL INSERT 72-4-0324 SCORPIOFLEX TOTAL KNEE CR 72-16-0512 72-16-0518 72-16-0912 72-16-1112 SERIES II TIBIAL BEAR INSERT NO 11 15MM 3052-1115 UHR BIPOLAR 26X52mm UH1-52-26 UHR BIPOLAR 26X58mm UH1-58-26 UHR BIPOLAR 28X48mm UH1-48-28 UHR BIPOLAR 28X52mm UH1-52-28 UHR BIPOLAR 28X55mm UH1-55-28 UHR BIPOLAR 32X49MM UH1-49-32 UHR BIPOLAR 32X51MM UH1-51-32 UHR BIPOLAR 32X52MM UH1-52-32 UHR BIPOLAR 32X54MM UH1-54-32 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Sold in
- Nationwide
- Made for
- Adults (18 years to 100 years)
This recall covers 12 products — see all of them.
Details
- Units affected
- 12,156 US
- Illnesses reported
- None reported
- Manufactured in
- 325 Corporate Dr, N/A, Mahwah, NJ, United States
- Classification
- Class II
- Recall number
- Z-3197-2026
In the FDA's words
- Description
- ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PATELLA 73-2710 SCORPIO INSET PATELLA 73-2610 SCORPIO M-DOME PATELLA 73-0110 73-0710 73-0910 SCORPIO RECESSED PATELLA 26MM 3044-0026 SCORPIO TS TIB INS 72-4-9114 SCORPIO TS TIBIAL INSERT 72-4-0324 SCORPIOFLEX TOTAL KNEE CR 72-16-0512 72-16-0518 72-16-0912 72-16-1112 SERIES II TIBIAL BEAR INSERT NO 11 15MM 3052-1115 UHR BIPOLAR 26X52mm UH1-52-26 UHR BIPOLAR 26X58mm UH1-58-26 UHR BIPOLAR 28X48mm UH1-48-28 UHR BIPOLAR 28X52mm UH1-52-28 UHR BIPOLAR 28X55mm UH1-55-28 UHR BIPOLAR 32X49MM UH1-49-32 UHR BIPOLAR 32X51MM UH1-51-32 UHR BIPOLAR 32X52MM UH1-52-32 UHR BIPOLAR 32X54MM UH1-54-32 Orthopaedic implant components manufactured from UHMWPE (polyethylene) Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. 69-2854 UDI-DI 07613327025491 Lots MLKY8A MLL8E1 82-2-0512 UDI-DI 04546540400611 Lot 23325151 82-2-0518 UDI-DI 04546540400635 Lot 23640051 82-2-0912 UDI-DI 04546540400758 Lot 26862851 82-3-0318 UDI-DI 04546540364319 Lot 28208551 78-5-0910 UDI-DI 07613327366006 Lots 30582401 30582601 73-2710 UDI-DI 07613327033847 Lot 14HDSA 73-2610 UDI-DI 07613327014082 Lots 193307A 193307E 73-0110 UDI-DI 07613327050288 Lot HHPW 73-0710 UDI-DI 07613327050295 Lots 8P83 DL4V 73-0910 UDI-DI 07613327050332 Lot 14NACD 3044-0026 UDI-DI 07613327033779 Lots 14SCVA 14SCVD 14SCVF 72-4-9114 UDI-DI 07613327034875 Lot 8H39KP 72-4-0324 UDI-DI 07613327034356 Lot 23505751 72-16-0512 UDI-DI 07613154020515 Lots 17376651 19093651 23125951 27835951 72-16-0518 UDI-DI 07613154020539 Lot 25621551 72-16-0912 UDI-DI 07613154020652 Lot 17879451 72-16-1112 UDI-DI 07613154020737 Lot 22764851 3052-1115 UDI-DI 07613327034240 Lot MJ7YY6 UH1-52-26 UDI-DI 07613327017168 Lot 5J8PM2 UH1-58-26 UDI-DI 07613327017359 Lots 1R8TNH 414A2L 4W1HD7 AV0J7J E226XA E50DAW M31665 WK1WP2 X90ML6 UH1-48-28 UDI-DI 07613327017014 Lots N046DW N06WWP N20WDM N75AY1 P419XW RP3D0X TD218R V15WA1 VH06PA VN3P18 VR853H WM8KJT XE2ND8 YH0N93 UH1-52-28 UDI-DI 07613327017182 Lot V02MME UH1-55-28 UDI-DI 07613327017281 Lot R30MME UH1-49-32 UDI-DI 07613327017069 Lot 324H9H UH1-51-32 UDI-DI 07613327017120 Lot 9J6N9E UH1-52-32 UDI-DI 07613327017212 Lots 5D4NMJ 9T612E UH1-54-32 UDI-DI 07613327017267 Lot MNNAAL Distribution: Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa. Affected quantity: 12,156 US
- Hazard
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
12 products in this recall
Howmedica Osteonics polyethylene implants recalled for wear risk
Howmedica Osteonics is recalling 12 orthopaedic implant components made from polyethylene material that was aged more than five years before use. The aged material may oxidize, which can increase wear and lead to wear debris, fractured or dislodged parts, and may require revision surgery. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)Z-3195-2026 · 12,156 US
- OMNIFIT SERIES I ACET. INSERT 10¿ 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured from UHMWPE (polyethylene)Z-3196-2026 · 12,156 US
- ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING P… (this page)Z-3197-2026 · 12,156 US
- TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured fr…Z-3198-2026 · 12,156 US
- ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-M071-6-700 ARC2F ACETABULAR…Z-3199-2026 · 12,156 US
- DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING G…Z-3200-2026 · 12,156 US
- ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-2246 ALL POLY CONSTRAINED INS…Z-3201-2026 · 12,156 US
- BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-…Z-3202-2026 · 12,156 US
- SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyeth…Z-3203-2026 · 12,156 US
- TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)Z-3204-2026 · 12,156 US
- SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 607…Z-3205-2026 · 12,156 US
- SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant compone…Z-3206-2026 · 12,156 US