Moderate riskFDA DeviceClass II

Howmedica Osteonics SCORPIO patella implants recalled for premature wear risk

Howmedica Osteonics manufactured SCORPIO recessed metal-backed patella implants using polyethylene material that was stored longer than five years before use. This aged material can develop oxidation, which may cause the implant to wear faster than expected, potentially leading to device failure, fractures, or loosening.

From the FDA notice · Read the official notice ↗ · All Howmedica Osteonics recalls

What to do

  1. If you received one of these implants, contact your surgeon or healthcare provider with your device lot number to determine if yours is affected.
  2. Your doctor may recommend imaging or monitoring to assess the implant's condition.
  3. Do not attempt to remove or replace the device yourself; only a qualified surgeon should manage any needed revision.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Sold in
Nationwide
Made for
Adults (18 years to 100 years)

This recall covers 12 products — see all of them.

Details

Units affected
12,156 US
Illnesses reported
None reported
Manufactured in
325 Corporate Dr, N/A, Mahwah, NJ, United States
Classification
Class II
Recall number
Z-3203-2026

In the FDA's words

Description
SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene) Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation. 3045-0026 No UDI Lots 12WAHA 12WAHB HOHE HOVX 3045-0028 No UDI Lots 13VAW 13VAWA 3045-0030 No UDI Lots BIYIE HNZY 3045-0032 No UDI Lots BIYIH HOYK Distribution: Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa. Affected quantity: 12,156 US
Hazard
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

12 products in this recall

Howmedica Osteonics polyethylene implants recalled for wear risk

Howmedica Osteonics is recalling 12 orthopaedic implant components made from polyethylene material that was aged more than five years before use. The aged material may oxidize, which can increase wear and lead to wear debris, fractured or dislodged parts, and may require revision surgery. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Howmedica Osteonics

See all

Related topics