Moderate riskFDA DeviceClass II

Alere, McKesson, Fisher Scientific, Cardinal Health, and Medline hCG rapid tests recalled for false positive results

Certain hCG rapid pregnancy tests may incorrectly detect low levels of hCG below their stated sensitivity range, potentially leading to false positive results. The affected tests were distributed nationwide and internationally and include multiple brand names sold under McKesson, Alere, Fisher Scientific, Cardinal Health, and Medline labels. No injuries have been reported.

From the FDA notice · Read the official notice ↗

What to do

  1. If you have one of the affected hCG test kits, do not use it and contact your healthcare provider if you have concerns about a pregnancy test result.
  2. Contact the manufacturer or distributor listed on your test kit packaging for instructions on obtaining a replacement or refund.
  3. If you received a positive result on one of these tests, confirm the result with a clinical blood test ordered by your doctor.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/Urine¿hCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG;
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 4 products — see all of them.

Details

Units affected
50,265,835
Illnesses reported
None reported
Manufactured in
9942 mesa rim rd, San Diego, CA, United States
Classification
Class II
Recall number
Z-3271-2026

In the FDA's words

Description
McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/Urine¿hCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG; hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result. All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 5001/ 20612479202744/1155958203, 12/15; 92210/10815845020130/1156015803, 07/16; 23900530/00613647000571/1155810105, 2018-01-15; 23900531/00613647000533/1155810105, 2018-01-15; 5002/20612479202768/1155958305, 12/15; 92215/10815845020123/1156015903, 07/16; B-1077-23/20885380020268/1155904805, 9/2014; MPH22025HCG/10884389191880/1156093402, 04-01-2015 Distribution: Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden. Affected quantity: 50,265,835
Hazard
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.

4 products in this recall

Alere hCG pregnancy tests recalled for false positive risk

Alere is recalling four groups of hCG rapid tests, which detect the pregnancy hormone hCG, because they may pick up very low hCG levels below the stated sensitivity. This can lead to a false positive result. No injuries have been reported, and the tests were sold across the U.S. and in several other countries.

The agency lists each product separately. Check every one you might have.