FDA-Device2026-10-07Class IIPROCESSING DEFECT

Philips Ingenia Elition MRI systems recalled for incorrect DTI scan exports

OTHERNationwide distributionMedical devices

What this means

Philips Ingenia Elition S and X MRI systems running software versions R11, R12, or R13 export brain diffusion tensor imaging (DTI) scans with incorrect directional data when using Enhanced MR DICOM format. This error can prevent third-party brain mapping software from correctly processing the scans.

  • Check your MRI system model and serial number against the recall list
  • Contact Philips North America for a software update or corrective action
  • Do not use Enhanced MR DICOM exports for DTI tractography until corrected
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.
Affected count
340
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Ingenia Elition S and Ingenia Elition X.; Model number (REF): 1. 781357 (S), 2. 782106 (S), 3. 782137 (S), 4. 782157 (S), 5. 781358 (X), 6. 782107 (X), 7. 782119 (X), 8. 782136 (X), 9. 782158 (X), 10. 782198 (X); Software Versions: R11, R12 and R13;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3311-2026