FDA-Device2026-10-07Class IIMISBRANDING

Percussionaire IPV-1C and Phasitron 5 ventilators need updated anesthetic safety warning

FALSE CLAIMSNationwide distributionMedical devices

What this means

The FDA has updated labeling for Percussionaire's IPV-1C and Phasitron 5 intrapulmonary percussive ventilators to warn against use with volatile anesthetic delivery systems. When used together, IPV may interfere with controlled anesthetic delivery, potentially causing unwanted changes in anesthetic depth and serious side effects including irregular heart rate, high brain pressure, or unintended awareness during surgery.

  • Review the updated instruction manual (P20045 Rev B) before using IPV devices in anesthesia settings
  • Do not combine IPV ventilation with volatile anesthetic delivery systems
  • Contact Percussionaire Corporation if you have questions about safe use
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Affected count
229,105
Manufactured in
130 Mcghee Rd Ste 109, Sandpoint, ID, United States
Products
Intrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3321-2026