FDA-Device2026-10-07Class IIMISBRANDING

Percussionaire Impulsator and Phasitron 5 IPV devices need updated anesthetic safety instructions

FALSE CLAIMSNationwide distributionMedical devices

What this means

Percussionaire has updated labeling for Impulsator and Phasitron 5 ventilation devices to warn against using them together with volatile anesthetic delivery systems. When used together, these devices may interfere with proper anesthetic delivery, potentially causing unintended changes in anesthetic depth, heart rate, blood pressure, or patient awareness during surgery.

  • Review the updated instruction manual (P20047 Rev E) before using these devices with anesthetic systems
  • Do not use Impulsator (S00010) or Phasitron 5 (P5) in conjunction with volatile anesthetic delivery systems
  • Contact Percussionaire Corporation or your device distributor if you have questions about safe use
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
Affected count
228,277
Manufactured in
130 Mcghee Rd Ste 109, Sandpoint, ID, United States
Products
Impulsator, Product Number: S00010, Phasitron 5, Product Number: P5 used with Instruction Manual

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3323-2026