- 2025-07-16FDA-DeviceClass IIVersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J RF WIRE 63CM/55DEG/D0) VXSK0103 VXSK (180P RF WIRE 63CM/45DEG/D0) VXSK0121 VXSK (180P RF WIRE 63CM/45DEG/D1) VXSK0122 VXSK (180P RF WIRE 63CM/55DEG/D0) VXSK0123 VXSK (180P RF WIRE 63CM/55DEG/D1) VXSK0124 VXSK (180P RF WIRE 63CM/90DEG/D0) VXSK0125 VXSK (230J RF WIRE 63CM/45DEG/D0) VXSK0111 VXSK (230J RF WIRE 63CM/45DE
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
- 2025-05-07FDA-DeviceClass IIIceSeed 1.5 CX NEEDLE OUS, Cryoablation Needle, REF H7493967534170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
- 2025-05-07FDA-DeviceClass IIIceSeed 1.5 CX NEEDLE US, Cryoablation Needle, REF H7493967433170. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
- 2025-05-07FDA-DeviceClass IIIceSeed 1.5 CX S NEEDLE OUS, Cryoablation Needle, REF H7493967334100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
- 2025-05-07FDA-DeviceClass IIIceSeed 1.5 CX S NEEDLE US, Cryoablation Needle, REF H7493967233100. The needles are intended to convert high-pressure gas to either a very cold Freezing application or to a warm Thawing application.
Boston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
- 2025-02-05FDA-DeviceClass IIHOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553570
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIHOT AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553560
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIHOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553550
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIHOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIHOT AXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553530
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIHOT AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553520
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Delivery System 15mmx10mm UPN: M00553750
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Delivery System 10mmx10mm UPN: M00553740
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553690
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553680
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553670
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553660
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-02-05FDA-DeviceClass IIAXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553640
Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
- 2025-01-22FDA-DeviceClass IVISIONIST CRT-P, EL (Model Number U225 and U226) and EL MRI (Model Number U228)
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- 2025-01-22FDA-DeviceClass IVALITUDE CRT-P, EL (Model Number U125) and EL MRI (Model Number U128)
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- 2025-01-22FDA-DeviceClass IALTRUA 2 DR Pacemaker, SL (Model Number S702) and EL (Model Number S722)
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- 2025-01-22FDA-DeviceClass IESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131)
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- 2025-01-22FDA-DeviceClass IPROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221), and EL MRI (Model Number L231).
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- 2025-01-22FDA-DeviceClass IACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- 2024-12-04FDA-DeviceClass IIAMS Artificial Urinary Sphincter Pressure Regulating Balloon 61-70 cm H2O
Potential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
- 2024-12-04FDA-DeviceClass IIAMS Artificial Urinary Sphincter Control Pump
Potential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
- 2024-11-20FDA-DeviceClass IBoston Scientific SMARTFREEZE CRYO CONSOLE: Console: Material Number M004CRBS4000; CONSOLE REFURB: Material Number M004CRBS400R0; CONSOLE HOSPITAL, Material Number M004CRBS400H0, CONSOLE ZERO, Material Number M004CRBS400Z0
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
- 2024-11-20FDA-DeviceClass IBoston Scientific POLARx FIT BALLOON CATHETER LT US, Material Number M004CRBS2160
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
- 2024-11-20FDA-DeviceClass IBoston Scientific POLARx FIT BALLOON CATHETER LT OUS, Material Number M004CRBS2110
Boston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.