Serious riskFDA DeviceClass II

Baxter TruSystem 7000 U operating table recalled for possible unexpected drop

The main lift on the TruSystem 7000 U operating table may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after a drop, and an uncontrolled drop may cause abrupt patient displacement during surgery.

From the FDA notice · Read the official notice ↗ · All Baxter recalls

What to do

  1. Set aside affected units and stop using them in procedures until you contact Baxter Healthcare Corporation.
  2. Check the serial numbers: 110188921, 110188922, and 110206662 (UDI-DI 00887761967519, product code 1723633).
  3. The notice doesn't say what remedy is available; contact Baxter Healthcare Corporation for instructions.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
TruSystem 7000 U Operating Table (dV), Product Code 1723633
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 2 products — see all of them.

Details

Units affected
62 units
Illnesses reported
None reported
Manufactured in
25212 W Il Route 120, Round Lake, Illinois, United States
Classification
Class II
Recall number
Z-0332-2027

In the FDA's words

Description
TruSystem 7000 U Operating Table (dV), Product Code 1723633 The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. UDI-DI: 00887761967519; Serial Number: 110188921, 110188922, and 110206662 Distribution: US Nationwide distribution in the states of IL, MO, TX. Affected quantity: 62 units
Hazard
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

2 products in this recall

Baxter TruSystem operating tables recalled for unexpected drop risk

Baxter is recalling two products: the TruSystem 7000 U Operating Table (dV), product code 1723633, and the mobile column TruSystem 7500 U, product code 1730720. The main lift may unexpectedly drop to its lowest position during use, and the tabletop may not be able to be raised again, which could cause abrupt patient displacement during surgery. The notice lists possible serious outcomes including traumatic injury and airway compromise. The products were distributed nationwide, with confirmed distribution in IL, MO and TX.

The agency lists each product separately. Check every one you might have.

More from Baxter

See all

Related topics