Baxter TruSystem 7500 U mobile column recalled for lift drop risk
Baxter is recalling one TruSystem 7500 U mobile column because its main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be raised again after a drop, and an uncontrolled drop could cause abrupt patient displacement during surgery, which may lead to traumatic injury, airway compromise, or surgical complications.
From the FDA notice · Read the official notice ↗ · All Baxter recalls
What to do
- Set the affected unit aside and do not use it for surgery.
- Confirm the unit by checking serial number 110183858 and UDI-DI 00887761974388.
- Contact Baxter Healthcare Corporation about the recall. The notice doesn't say how to get a repair or replacement.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Mobile column TruSystem 7500 U, Product Code 1730720
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 2 products — see all of them.
Details
- Units affected
- 1 unit
- Illnesses reported
- None reported
- Manufactured in
- 25212 W Il Route 120, Round Lake, Illinois, United States
- Classification
- Class II
- Recall number
- Z-0333-2027
In the FDA's words
- Description
- Mobile column TruSystem 7500 U, Product Code 1730720 The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. UDI-DI: 00887761974388; Serial number: 110183858 Distribution: US Nationwide distribution in the states of IL, MO, TX. Affected quantity: 1 unit
- Hazard
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
2 products in this recall
Baxter TruSystem operating tables recalled for unexpected drop risk
Baxter is recalling two products: the TruSystem 7000 U Operating Table (dV), product code 1723633, and the mobile column TruSystem 7500 U, product code 1730720. The main lift may unexpectedly drop to its lowest position during use, and the tabletop may not be able to be raised again, which could cause abrupt patient displacement during surgery. The notice lists possible serious outcomes including traumatic injury and airway compromise. The products were distributed nationwide, with confirmed distribution in IL, MO and TX.
The agency lists each product separately. Check every one you might have.
- TruSystem 7000 U Operating Table (dV), Product Code 1723633Z-0332-2027 · 62 units
- Mobile column TruSystem 7500 U, Product Code 1730720 (this page)Z-0333-2027 · 1 unit
More from Baxter
See all- Baxter TruSystem 7500 surgical table recalled for unexpected drop risk
SeriousFDANov 3, 2026
- Baxter VOLARA System recalled for possible nebulizer flow rate error
ModerateFDASep 30, 2026
- Baxter Dextrose Injection 70% recalled for stainless steel particles
SeriousFDASep 16, 2026