Serious riskFDA DeviceClass I

Intersurgical 22mm breathing circuits recalled for disconnection risk

The recall covers four Intersurgical 22mm Smoothbore breathing system models. The soft end connectors may disconnect from the tube, which could cause loss of ventilation or compromised ventilation. The FDA has classified this as a Class I recall, its most serious category.

From the FDA notice · Read the official notice ↗

What to do

  1. Contact the manufacturer and your care team before changing or stopping use of these breathing systems.
  2. Check whether your product has REF number 5018000, 5804000, 5805000, or 5805030.
  3. The notice doesn't say how to get a repair or replacement; ask Pulmodyne Inc. or your care provider.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Intersurgical Smoothbore Breathing Systems, Model Numbers: 1) REF: 5018000, 22mm Smoothbore tube with soft end connectors, 22F-22F, e 1.8m; 2) REF: 5804000, 22mm Smoothbore bilevel breathing system with swivel CO¿ leak port, e 1.8m; 3) REF: 5805000, 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO¿ leak port, e1.6m; 4) REF: 5805030, 22mm Smoothbore bilevel breathing system with swivel CO¿ leak port and Clear-Guard Midi filter e1.8m; breathing circuit
Sold in
US Distribution: FL, OH, MN, TX, VA, NY, NJ, CA, CO, IN.
Made for
All ages (under 100 years)

This recall covers 3 products — see all of them.

Details

Units affected
28368 units
Illnesses reported
None reported
Manufactured in
2055 Executive Dr, Indianapolis, Indiana, United States
Classification
Class I
Recall number
Z-0493-2027

In the FDA's words

Description
Intersurgical Smoothbore Breathing Systems, Model Numbers: 1) REF: 5018000, 22mm Smoothbore tube with soft end connectors, 22F-22F, e 1.8m; 2) REF: 5804000, 22mm Smoothbore bilevel breathing system with swivel CO¿ leak port, e 1.8m; 3) REF: 5805000, 22mm Smoothbore bilevel breathing system with monitoring line and swivel CO¿ leak port, e1.6m; 4) REF: 5805030, 22mm Smoothbore bilevel breathing system with swivel CO¿ leak port and Clear-Guard Midi filter e1.8m; breathing circuit Soft connectors may disconnect from the tube resulting in a loss of ventilation or compromised ventilation. no lot code Distribution: US Distribution: FL, OH, MN, TX, VA, NY, NJ, CA, CO, IN. Affected quantity: 28368 units
Hazard
Soft connectors may disconnect from the tube resulting in a loss of ventilation or compromised ventilation.

3 products in this recall

Intersurgical breathing systems recalled for possible connector disconnection

Pulmodyne is recalling 3 Intersurgical Smoothbore breathing system product groups, about 30,800 units in total, sold in FL, OH, MN, TX, VA, NY, NJ, CA, CO and IN. The soft connectors may disconnect from the tube, which could cause loss of ventilation or compromised ventilation. The notice reports no injuries.

The agency lists each product separately. Check every one you might have.