Intersurgical breathing systems recalled for possible disconnection
Pulmodyne Inc. is recalling Intersurgical Smoothbore Breathing Systems, models REF 5183000 and REF 5191000. The soft connectors may disconnect from the tube, which could cause loss of ventilation or compromised ventilation. The recall covers 1,975 units, and no injuries have been reported.
From the FDA notice · Read the official notice ↗
What to do
- Remove affected units from use, and set aside any product from the 11 listed lots, which have expiration dates from 10/17/2026 to 10/28/2026.
- Check lot numbers and model REFs (5183000 or 5191000) against the lot list in the notice.
- The notice doesn't say how to get a replacement or repair; contact Pulmodyne Inc. for next steps.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Intersurgical Smoothbore Breathing Systems, Model Numbers: 1) REF: 5183000, 22mm Smoothbore breathing system with monitoring line and exhalation valve, e 1.6m; 2) REF: 5191000, 22mm Smoothbore breathing system with monitoring line and exhalation valve, single limb, 1.8m; non-breathing valve
- Sold in
- US Distribution: FL, OH, MN, TX, VA, NY, NJ, CA, CO, IN.
- Made for
- All ages (under 100 years)
This recall covers 3 products — see all of them.
Details
- Units affected
- 1975 units
- Illnesses reported
- None reported
- Manufactured in
- 2055 Executive Dr, Indianapolis, Indiana, United States
- Classification
- Class I
- Recall number
- Z-0494-2027
In the FDA's words
- Description
- Intersurgical Smoothbore Breathing Systems, Model Numbers: 1) REF: 5183000, 22mm Smoothbore breathing system with monitoring line and exhalation valve, e 1.6m; 2) REF: 5191000, 22mm Smoothbore breathing system with monitoring line and exhalation valve, single limb, 1.8m; non-breathing valve Soft connectors may disconnect from the tube resulting in a loss of ventilation or compromised ventilation. Lot: 122-08976163, Exp Date: 10/17/2026; 122-08977504, Exp Date: 10/19/2026; 122-08998943, Exp Date: 10/23/2026; 122-09005722, Exp Date: 10/23/2026; 122-09007780, Exp Date: 10/24/2026; 122-09008574, Exp Date: 10/26/2026; 122-09011368, Exp Date: 10/26/2026; 122-09013637, Exp Date: 10/26/2026; 122-09014200, Exp Date: 10/27/2026; 122-09021717, Exp Date: 10/28/2026; 122-09021741, Exp Date: 10/28/2026 Distribution: US Distribution: FL, OH, MN, TX, VA, NY, NJ, CA, CO, IN. Affected quantity: 1975 units
- Hazard
- Soft connectors may disconnect from the tube resulting in a loss of ventilation or compromised ventilation.
3 products in this recall
Intersurgical breathing systems recalled for possible connector disconnection
Pulmodyne is recalling 3 Intersurgical Smoothbore breathing system product groups, about 30,800 units in total, sold in FL, OH, MN, TX, VA, NY, NJ, CA, CO and IN. The soft connectors may disconnect from the tube, which could cause loss of ventilation or compromised ventilation. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- Intersurgical Smoothbore Breathing Systems, Model Numbers:
1) REF: 5025000, 22mm Smoothbore single heated wire breathing system with water trap, e 1.6m;…Z-0492-2027 · 473 units
- Intersurgical Smoothbore Breathing Systems, Model Numbers:
1) REF: 5018000, 22mm Smoothbore tube with soft end connectors, 22F-22F, e 1.8m;
2) REF: 5804…Z-0493-2027 · 28368 units
- Intersurgical Smoothbore Breathing Systems, Model Numbers:
1) REF: 5183000, 22mm Smoothbore breathing system with monitoring line and exhalation valve, e… (this page)Z-0494-2027 · 1975 units