Serious riskFDA DeviceClass II

Boston Scientific OptiCross 6 catheter recalled for air embolism risk

Boston Scientific is updating the instructions for use (IFU) for its OptiCross 6 coronary imaging catheters after reports of air embolism events. The reports involved repeated flushing to restore image quality while the catheter was positioned in the patient. Intravascular air may cause an air embolism, with consequences that can include arrhythmia, myocardial infarction, cardiac arrest, and other serious injury.

From the FDA notice · Read the official notice ↗ · All Boston Scientific recalls

What to do

  1. Review the updated instructions for use from Boston Scientific before using these catheters.
  2. Avoid repeated flushing to restore image quality while the catheter is positioned in the patient.
  3. Check lot numbers: 122-08977093 (exp 10/20/2026), 122-08994147 (exp 10/22/2026), and 122-09014193.
  4. Check material numbers H7495181260, H7495181160, H7495181360, and H7495181460.
  5. The notice doesn't specify a replacement or removal remedy; contact Boston Scientific for guidance on affected units.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
OptiCross 6 Coronary Imaging Catheter, Material Numbers/UPN: H7495181260, H7495181160, H7495181360, H7495181460
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 6 products — see all of them.

Details

Units affected
1,288,662 total units
Illnesses reported
None reported
Manufactured in
1 Scimed Pl, Maple Grove, Minnesota, United States
Classification
Class II
Recall number
Z-0509-2027

In the FDA's words

Description
OptiCross 6 Coronary Imaging Catheter, Material Numbers/UPN: H7495181260, H7495181160, H7495181360, H7495181460 IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury. 1. Model Number: 722079. UDI Numbers: (01)00884838085268(21)1023, (01)00884838085268(21)646, (01)00884838085268(21)703904, (01)00884838085268(21)890, (01)00884838085268(21)1137, (01)00884838085268(21)1048, (01)00884838085268(21)401, (01)00884838085268(21)536, (01)00884838085268(21)1054, (01)00884838085268(21)1010, (01)00884838085268(21)266, (01)00884838085268(21)617, (01)00884838085268(21)246, (01)00884838085268(21)905, (01)00884838085268(21)809, (01)00884838085268(21)848, (01)00884838085268(21)1033, (01)00884838085268(21)836, (01)00884838085268(21)466, (01)00884838085268(21)1155, (01)00884838085268(21)673, (01)00884838085268(21)564, (01)00884838085268(21)1032, (01)00884838085268(21)1117, (01)00884838085268(21)820, (01)00884838085268(21)1211, (01)00884838085268(21)1213, (01)00884838085268(21)47, (01)00884838085268(21)1365, (01)00884838085268(21)881, (01)00884838085268(21)774, (01)00884838085268(21)955, (01)00884838085268(21)633, (01)00884838085268(21)703540, (01)00884838085268(21)296, (01)00884838085268(21)574, (01)00884838085268(21)842, (01)00884838085268(21)884, (01)00884838085268(21)991, (01)00884838085268(21)703336, (01)00884838085268(21)753, (01)00884838085268(21)703493, (01)00884838085268(21)790, (01)00884838085268(21)481, (01)00884838085268(21)407, (01)00884838085268(21)408, (01)00884838085268(21)877, (01)00884838085268(21)703395, (01)00884838085268(21)703323, (01)00884838085268(21)483, (01)00884838085268(21)241, (01)00884838085268(21)1108, (01)00884838085268(21)1363, (01)00884838085268(21)1414, (01)00884838085268(21)703404, (01)00884838085268(21)703387, (01)00884838085268(21)1355, (01)00884838085268(21)863, (01)00884838085268(21)350, (01)00884838085268(21)908, (01)00884838085268(21)1378, (01)00884838085268(21)262, (01)00884838085268(21)598, (01)00884838085268(21)275, (01)00884838085268(21)520, (01)00884838085268(21)593, (01)00884838085268(21)1018, (01)00884838085268(21)1432, (01)00884838085268(21)594, (01)00884838085268(21)999, (01)00884838085268(21)361, (01)00884838085268(21)526, (01)00884838085268(21)1021, (01)00884838085268(21)608, (01)00884838085268(21)611, (01)00884838085268(21)413, (01)00884838085268(21)895, (01)00884838085268(21)428, (01)00884838085268(21)883, (01)00884838085268(21)758, (01)00884838085268(21)274, (01)00884838085268(21)525, (01)00884838085268(21)885, (01)00884838085268(21)843, (01)00884838085268(21)754, (01)00884838085268(21)703528, (01)00884838085268(21)599, (01)00884838085268(21)1171, (01)00884838085268(21)817, (01)00884838085268(21)766, (01)00884838085268(21)878, (01)00884838085268(21)1170, (01)00884838085268(21)703310, (01)00884838085268(21)944, (01)00884838085268(21)703305, (01)00884838085268(21)1285, (01)00884838085268(21)854, (01)00884838085268(21)730, (01)00884838085268(21)703497, (01)00884838085268(21)957, (01)00884838085268(21)658, (01)00884838085268(21)1022, (01)00884838085268(21)621, (01)00884838085268(21)628, (01)00884838085268(21)795, (01)00884838085268(21)1007, (01)00884838085268(21)1289, (01)00884838085268(21)605, (01)00884838085268(21)65, (01)00884838085268(21)34, (01)00884838085268(21)1222, (01)00884838085268(21)769, (01)00884838085268(21)938, (01)00884838085268(21)655, (01)00884838085268(21)1311, (01)00884838085268(21)538, (01)00884838085268(21)989, (01)00884838085268(21)946, (01)00884838085268(21)888, (01)00884838085268(21)402, (01)00884838085268(21)1042, (01)00884838085268(21)374, (01)00884838085268(21)459, (01)00884838085268(21)1282, (01)00884838085268(21)841, (01)00884838085268(21)703532, (01)00884838085268(21)1360, (01)00884838085268(21)1359, (01)00884838085268(21)1358, (01)00884838085268(21)50, (01)00884838085268(21)326, (01)00884838085268(21)216, (01)00884838085268(21)927, (01)00884838085268(21)981, (01)00884838085268(21)923, (01)00884838085268(21)1163, (01)00884838085268(21)416, (01)00884838085268(21)424, (01)00884838085268(21)802, (01)00884838085268(21)751, (01)00884838085268(21)436, (01)00884838085268(21)910, (01)00884838085268(21)1242, (01)00884838085268(21)251, (01)00884838085268(21)1156, (01)00884838085268(21)1049, (01)00884838085268(21)1103, (01)00884838085268(21)1361, (01)00884838085268(21)1041, (01)00884838085268(21)29, (01)00884838085268(21)930, (01)00884838085268(21)228, (01)00884838085268(21)30, (01)00884838085268(21)366, (01)00884838085268(21)543, (01)00884838085268(21)921, (01)00884838085268(21)1200, (01)00884838085268(21)359, (01)00884838085268(21)355, (01)00884838085268(21)357, (01)00884838085268(21)356, (01)00884838085268(21)1090, (01)
Hazard
IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury.

6 products in this recall

Boston Scientific OptiCross catheters need new instructions for air embolism

Boston Scientific is updating the instructions for use for six OptiCross coronary and peripheral imaging catheter lines, covering about 1.29 million units. The update follows reports of air embolism events and of repeated flushing to restore image quality while the catheter was inside the patient. The agency says intravascular air may cause air embolism, with effects of varying severity such as heart rhythm and blood flow problems, heart attack, cardiac arrest, and other serious injury. No injuries are reported in the notice.

The agency lists each product separately. Check every one you might have.