Serious riskFDA DeviceClass II

Boston Scientific OptiCross 35 catheter updated for air embolism risk

Boston Scientific updated the instructions for use for its OptiCross 35 Peripheral Imaging Catheter after reports of air embolism events. Some users flushed the catheter repeatedly to restore image quality while it was inside the patient. Intravascular air may cause an air embolism, which can lead to serious harm including cardiac arrest.

From the FDA notice · Read the official notice ↗ · All Boston Scientific recalls

What to do

  1. Follow the updated instructions for use when using these catheters.
  2. Check the material numbers H7493932800350, H7493932801350, and H7493932802350 to identify affected catheters.
  3. The notice doesn't say how to get a replacement; contact Boston Scientific Corporation with questions.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
OptiCross 35 Peripheral Imaging Catheter, Material Numbers/UPN: H7493932800350, H7493932801350, H7493932802350
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 6 products — see all of them.

Details

Units affected
1,288,662 total units
Illnesses reported
None reported
Manufactured in
1 Scimed Pl, Maple Grove, Minnesota, United States
Classification
Class II
Recall number
Z-0513-2027

In the FDA's words

Description
OptiCross 35 Peripheral Imaging Catheter, Material Numbers/UPN: H7493932800350, H7493932801350, H7493932802350 IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury. 1. Model No. EG38-J10UT; UDI-DI: 04961333239474, 04961333239481, 04961333239498, 04961333239504, 04961333239528, 04961333252770; All Serial Numbers. 2. Model No. EG34-J10U; UDI-DI: 04961333239672, 04961333239689, 04961333239696, 04961333239702, 04961333239726, 04961333248834, 04961333248841, 04961333248858, 04961333248872, 04961333248896, 04961333252732; All Serial Numbers. Distribution: Worldwide distribution - US Nationwide and Territories and the countries of EMEA, Canada, Japan, LATAM and Asia Pacific countries/regions. Affected quantity: 1,288,662 total units
Hazard
IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury.

6 products in this recall

Boston Scientific OptiCross catheters need new instructions for air embolism

Boston Scientific is updating the instructions for use for six OptiCross coronary and peripheral imaging catheter lines, covering about 1.29 million units. The update follows reports of air embolism events and of repeated flushing to restore image quality while the catheter was inside the patient. The agency says intravascular air may cause air embolism, with effects of varying severity such as heart rhythm and blood flow problems, heart attack, cardiac arrest, and other serious injury. No injuries are reported in the notice.

The agency lists each product separately. Check every one you might have.