Moderate riskFDA DeviceClass II

Boston Scientific OptiCross 6 HD catheter recalled for air embolism risk

Boston Scientific updated the instructions for use for its OptiCross 6 HD coronary imaging catheter after reports of air embolism events. Some of these followed repeated flushing to restore image quality while the catheter was inside the patient. The company says intravascular air may cause an air embolism, with effects ranging in severity from heart rhythm changes and reduced blood flow to heart attack, cardiac arrest, or other serious injury.

From the FDA notice · Read the official notice ↗ · All Boston Scientific recalls

What to do

  1. Review the updated instructions for use from Boston Scientific before using these catheters.
  2. Check whether the catheter is one of these material numbers: H74939354050, H74939354060, H74939354080, H74939354090, or H74939354100.
  3. Check the lot and expiration date: lot 122-08998951, expiration 10/23/2026.
  4. The notice doesn't specify a remedy; contact Boston Scientific for the updated instructions and next steps.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
OptiCross 6 HD Coronary Imaging Catheter, Material Numbers/UPN: H74939354050, H74939354060, H74939354080, H74939354090, H74939354100
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 6 products — see all of them.

Details

Units affected
1,288,662 total units
Illnesses reported
None reported
Manufactured in
1 Scimed Pl, Maple Grove, Minnesota, United States
Classification
Class II
Recall number
Z-0511-2027

In the FDA's words

Description
OptiCross 6 HD Coronary Imaging Catheter, Material Numbers/UPN: H74939354050, H74939354060, H74939354080, H74939354090, H74939354100 IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury. CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 584713; CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 584713; CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 584713; CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 584713; CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 585725; CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 585725; CATALOG #MA1DH7533C, GTIN 10888439997057 (ea) 50888439997055 (cs), LOT # 585725; CATALOG #MAT4226716, GTIN 10888439989779 (ea) 50888439989777 (cs), LOT # 554018; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #MAT4226780, GTIN 10888439986112 (ea) 50888439986110 (cs), LOT # 551641; CATALOG #OVCVM01028, GTIN 10888439986006 (ea) 50888439986004 (cs), LOT # 551639; CATALOG #OVCVM01028, GTIN 10888439986006 (ea) 50888439986004 (cs), LOT # 551639; CATALOG #OVCVM01028, GTIN 10888439986006 (ea) 50888439986004 (cs), LOT # 551639; CATALOG #OVCVM01028, GTIN 10888439986006 (ea) 50888439986004 (cs), LOT # 551639; CATALOG #OVCVM01028, GTIN 10888439986006 (ea) 50888439986004 (cs), LOT # 551639; CATALOG #MAT4226715, GTIN 10888439976380 (ea) 50888439976388 (cs), LOT # 543225; CATALOG #MAT4226715, GTIN 10888439976380 (ea) 50888439976388 (cs), LOT # 543225; CATALOG #OVCVM01027, GTIN 10888439972450 (ea) 50888439972458 (cs), LOT # 541281; CATALOG #MAT4226714, GTIN 10888439955798 (ea) 50888439955796 (cs), LOT # 528088; CATALOG #MAT4226714, GTIN 10888439955798 (ea) 50888439955796 (cs), LOT # 528088; CATALOG #MAT4226714, GTIN 10888439955798 (ea) 50888439955796 (cs), LOT # 528088; CATALOG #MAT4226714, GTIN 10888439955798 (ea) 50888439955796 (cs), LOT # 528088; CATALOG #OROMH47346, GTIN 10888439952490 (ea) 50888439952498 (cs), LOT # 520229; CATALOG #OROMH47346, GTIN 10888439952490 (ea) 50888439952498 (cs), LOT # 520229; CATALOG #OVCVM01026, GTIN 10888439917796 (ea) 50888439917794 (cs), LOT # 502500; CATALOG #OVCVM01026, GTIN 10888439917796 (ea) 50888439917794 (cs), LOT # 502500; CATALOG #MAT4226779, GTIN 10888439913897 (ea) 50888439913895 (cs), LOT # 497830; CATALOG #MAT4226713, GTIN 10888439907803 (ea) 50888439907801 (cs), LOT # 497831; CATALOG #MAT4226713, GTIN 10888439907803 (ea) 50888439907801 (cs), LOT # 497831; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 609466; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 609466; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 609466; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 609466; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 609466; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 609466; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 654186; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 654186; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 654186; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 654186; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 703762; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 734712; CATALOG #MAT422677D, GTIN 10888439897029 (ea) 50888439897027 (cs), LOT # 734712; CATALOG #MAT422677D, GTIN 10888439897029 (
Hazard
IFU updates in response to reports of air embolism events and reports of repeated flushing to restore image quality while the catheter was positioned within the patient. Intravascular air may cause air embolism, resulting in clinical consequences of varying severity, including ST-segment changes, reduced blood flow, arrhythmia, myocardial infarction, cardiac arrest, and other serious injury.

6 products in this recall

Boston Scientific OptiCross catheters need new instructions for air embolism

Boston Scientific is updating the instructions for use for six OptiCross coronary and peripheral imaging catheter lines, covering about 1.29 million units. The update follows reports of air embolism events and of repeated flushing to restore image quality while the catheter was inside the patient. The agency says intravascular air may cause air embolism, with effects of varying severity such as heart rhythm and blood flow problems, heart attack, cardiac arrest, and other serious injury. No injuries are reported in the notice.

The agency lists each product separately. Check every one you might have.

More from Boston Scientific