FDA-Device2025-12-24Class IIPROCESSING DEFECT

Intelerad InteleShare software recall for measurement inaccuracy in medical imaging

Check Intelerad InteleShare software version

Intelerad InteleShare medical imaging software versions 3.24.3 through 3.25.2 contain a bug that can produce inaccurate measurement values when certain viewing features are used together. This could lead to incorrect measurements on medical images, potentially affecting diagnosis and treatment planning.

  • Check if your facility uses Intelerad InteleShare software versions 3.24.3–3.25.2
  • If affected, contact Intelerad for a software update or patch
  • Review recent measurements taken with multiplanar reconstruction and manual rotation enabled for accuracy
Hazard

Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.

Sold states
US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE
Affected count
924
Manufactured in
305 Church At North Hills St, Raleigh, NC, United States
Products
Intelerad InteleShare software, with ProViewer component

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0941-2026