FDA-Device2026-04-08Class IIPROCESSING DEFECT
Intuitive Surgical da Vinci S, Si Double Fenestrated Grasper recalled for broken grip cables
OTHERNationwide distribution
Stop using affected da Vinci grasper instruments
Intuitive Surgical has recalled certain da Vinci S and Si Double Fenestrated Grasper reusable surgical instruments due to reports of broken or frayed grip cables. The cables transmit motion from the surgeon's input to the instrument tip inside the patient. A cable failure during surgery could compromise the surgeon's ability to control the instrument.
- Check if your facility has any affected instruments by matching the batch/lot numbers listed in the recall notice
- Stop using any graspers matching the recalled batch numbers immediately
- Contact Intuitive Surgical for instructions on inspection, repair, or replacement
- Review affected instruments for cable damage before any further use
Hazard
Due to increased complaints for broken/frayed grip cables for reusable instruments.
Sold states
Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Affected count
2370
Manufactured in
1266 Kifer Rd, Sunnyvale, CA, United States
Products
Brand Name: da Vinci S, Si Double Fenestrated Grasper Instrument Product Name: da Vinci S, Si Double Fenestrated Grasper Instrument Model/Catalog Number: 420189 Software Version: NA Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: NA
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1697-2026More Intuitive Surgical, Inc. recalls
- FDA-Device2026-04-08Intuitive Surgical da Vinci S/Si Cautery Hook instruments recalled for frayed or broken cables
- FDA-Device2026-04-08Intuitive Surgical da Vinci S/Si Mega Needle Driver recalled for broken grip cables
- FDA-Device2026-04-08Intuitive Surgical da Vinci S/Si Grasping Retractor recalled for broken grip cables
- FDA-Device2026-04-08Intuitive Surgical da Vinci S and Si Monopolar Curved Scissors recalled for broken grip cables
- FDA-Device2026-04-08Intuitive Surgical da Vinci Tenaculum Forceps recalled for frayed or broken pitch cables
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief