FDA-Device2026-04-08Class IIPROCESSING DEFECT
Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTO
Hazard
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.
Intended for
Adults
Illnesses reported
None reported
Sold states
US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.
Affected count
14,379 kits total
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Products
Medline kits containing Olympus biopsy valves: 1. BRONCH PROCEDURE KIT, DYKE2091 2. BRONCHSCOPY, DYNJ900898G DYNJ901922I 3. BRONCHSCOPY PACK, DYNJ38313B 4. BRONCHOSCOPY PACK 0120367-LF, DYNJ32750G DYNJ32750I 5. KIT, WR THORACIC, DYNJ911252 6. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146B 7. LEX THORACIC, DYNJ902016K DYNJ902016L 8. ORGAN PROCUREMENT, DYNJ908686 9. PULMONARY PACK, DYKE1859 10. THORACIC, DYNJ901666L 11. THORACIC ROBOTICS, DYNJ908777A 12. THORACOTOMY/THORACOSCOPY, DYNJ900482N DYNJ900482O 13. XPS, DYNJ907605
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1710-2026More Medline recalls
- FDA-Device2026-09-09Medline Convenience Kits recalled for potential sterility breach in antiseptic applicators
- FDA-Device2026-09-09Medline Convenience Kits with BD ChloraPrep Clear applicators recalled for contamination risk
- FDA-Device2026-09-09Medline Convenience Kits recalled for sterility packaging breach risk
- FDA-Device2026-09-09Medline Convenience Kits with BD ChloraPrep Clear applicators recalled for defective packaging seals
- FDA-Device2026-09-09Medline Arterial Line Insertion Kit recalled for sterility packaging breach risk
Want us to check the next one for you?
We read every CPSC, FDA, USDA, and NHTSA notice and email you a plain-English Sunday brief covering your household — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.