Moderate riskFDA DeviceClass II

Civco eTRAX Needle System Starter Kit recalled for inaccurate needle position display

Civco Medical Instruments Co. Inc. is recalling certain eTRAX Needle System Starter Kits (18G) used with Aurora Trackers due to a manufacturing error. The needle sensor may not correctly show the needle tip position on the display, which could affect proper use of the device.

From the FDA notice · Read the official notice ↗

What to do

  1. Check if you have lot number A233347, A228677, A268116, A268238, A272513, A279457, A280388, or A281840
  2. Stop using the affected eTRAX Needle System Starter Kit immediately
  3. Contact Civco Medical Instruments for instructions on how to return or replace your device

Is yours affected?

Compare your label or packaging with the recalled product.

Product
eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152
Sold in
US States: OH, PA, WA. China.
Made for
All ages (0 months and older)

This recall covers 8 products — see all of them.

Details

Units affected
20 units
Illnesses reported
None reported
Manufactured in
102 1st St S, Kalona, IA, United States
Classification
Class II
Recall number
Z-1734-2026

In the FDA's words

Description
eTRAX Needle System Starter Kit 18G (for Aurora Trackers), Part Number 667-152 There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. UDI-DI: 00841436111065; Lot Number: A233347, A228677, A268116, A268238, A272513, A279457, A280388 , A281840 Distribution: US States: OH, PA, WA. China. Affected quantity: 20 units
Hazard
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

8 products in this recall

Civco eTRAX needle systems recalled for tip position error

Civco is recalling 8 eTRAX needle system starter kits and needle sensors used with Aurora trackers. An error in inspection and programming may cause the needle tip position to be shown incorrectly on the user interface. No injuries have been reported.

The agency lists each product separately. Check every one you might have.